Troubleshootimg
46
Hei-VAP Precision
Error / message
Cause
The temperature sensor
for calculating the vapor
temperature is not
connected, defective or the
temperature is above the
measured range.
▪ Vapor sensor broken
▪ not connected
▪ above the temp.
range
➜
Let the boiling temperature
sensor cool
➜
Connect the vapor
temperature sensor
➜
Perform visual check
➜
Contact service
The vapor temperature
sensor is defective or the
temperature is below the
measured range
▪ Short circuit of the
vapor sensor
▪ below the temp.
range
➜
Heat the boiling
temperature sensor
➜
Perform visual check
➜
Contact service
The Autoaccurate sensor is
defective or the tempera-
ture is below the measured
range.
▪ Autoaccurate sensor
short circuit
▪ below the temp.
range
➜
Heat the Autoaccurate
sensor
➜
Perform visual check
➜
Contact service
The Autoaccurate sensor is
defective or the tempera-
ture is above the measured
range.
▪ Autoaccurate sensor
is broken
▪ not connected
▪ above the temp.
range
➜
Let the Autoaccurate
sensor cool
➜
Connect the Autoaccurate
sensor
➜
Perform visual check
➜
Contact service
The transportation protection
device is active, please raise
the lift and briefly switch the
unit off and then on again.
The transportation
protection device is still
activated.
➜
Please raise the lift and
briefly switch the unit off
and then on again.
An error occurred in the data
bus.
Cable connection
between the control
panel and the base unit
was disconnected. An
error occurred in the
data bus.
➜
Restore the cable
connection between the
control panel and the base
unit. Switch the power
switch off and on again!
➜
Contact service
The overheating protection
was triggered or the heater
is defectivet
Heater is defective, i.e.
the overheating protec-
tion was triggered.
➜
If the medium was in
the heating bath: Let the
heating bath cool and reset
the maximum temperature
limiter (see chapter
“Cleaning and mainte-
nance, Maintenance”)
➜
If the medium was not in the
heating bath: Contact service
A temperature above the
safety temperature was
measured in the heating
bath. The temperature
measurement is defective or
the medium filled in was too
hot.
Heating bath safety
temperature is
exceeded
➜
Switch off units
➜
Cool the heating bath
medium
➜
Contact Service if media
was not filled in