background image

Non-linear Error Compensation

To work with the non-linear error compensation it is necessary to

activate the function via the operating parameter P40.

traverse the reference marks after switching on.

enter the compensation values in the table.

For every axis compensation values can be entered over 16 compensation points.
To determine the compensation values with a comparator system from
HEIDENHAIN, such as VM 101, you must select the REF display.

Selecting the Compensation Value Table

Select the operating parameter  P00 and enter the code number  105 296. Use
the following keys for the entries:

Key

Function

MOD

Save input value and select next input parameter.

   1 /    2

Save input value and select preceding input value.

SPEC FCT

Select REF display.

ENT

Save entry.

Exit compensation value table.

CL

Delete entry.

Delete all compensation values.

Enter the parameters and compensation values as follows:

Display

Entry

Enter the axis to be compensated, e.g. X.

Enter the axis causing the error, e.g. X, i.e. X = F(X).

Enter the datum on the axis causing the error.

Enter the distance of the compensation points on the error-
causing axis, e.g. 14 (= 2

14

 µm = 16.384 mm).

Minimum input value: 10 (= 1.024 mm)
Maximum input value: 23 (= 8388.608 mm)

Select compensation point No. 1. The compensation point
number can be seen while pressing the MOD key. After letting
go of the MOD key the coordinates of the selected compensation
point can be seen in the upper line. Enter the compensation value
in the lower line.

Enter all following compensation points.

Delete all compensation values:

Display

Entry

Press key CL.

Press key ENT. Compensation values are deleted.

Содержание ND 510

Страница 1: ...point 2 currently active SCL Scaling factor active Define workpiece edge as datum Blinking Waiting for operator to confirm selection Define centerline between two workpiece edges as datum Blinking Wai...

Страница 2: ...ay shows the value last assigned to the reference mark position the REF indicator glows and the decimal point blinks Cross over the reference mark in each axis Move the axes one after the other until...

Страница 3: ...sed on the new datum This function ends automatically Centerline between two workpiece edges as datum Select the datum Press the SPEC FCT key twice The indicator Centerline as datum starts blinking Pr...

Страница 4: ...nput value SPEC FCT Select REF display ENT Save entry Exit compensation value table CL Delete entry Delete all compensation values Enter the parameters and compensation values as follows Display Entry...

Страница 5: ...rect input value Input frequency too high for encoder input will occur for example when traverse speed too high Internal counter overflow Error while crossing over reference marks To clear the error m...

Страница 6: ...ber Protected Operating Parameters In order to change protected operating parameters the code number 95 148 must be entered via P00 Code They remain accessible until the position display is switched o...

Страница 7: ...with 2000 SP Distance coded with 5000 SP Encoder Monitoring off Alarm Off Encoder monitoring A Monitoring on Alarm On Axis display A Display measured position Axis Do not display measured position no...

Страница 8: ...le before opening the housing Connect a protective ground This connection should never be interrupted Danger to internal components Do not engage or disengage any connections while the unit is under p...

Отзывы: