Heathrow Scientific HS28271 Скачать руководство пользователя страница 4

PART NUMBER

DESCRIPTION

PLUG TYPE

HS28270

220/230 VAC

U.K.

HS28271

220/230 VAC

Cont. Europe/Israel

HS28272

110/120 VAC

U.S./Taiwan/Japan

HS28280

Replacement Part Kit

 —

2 extra blades, 2 extra heating elements and 2 extra PTFE cloths

Parts Guide

Maintenance 

 

  1.

 Unplug the unit.

    

    2.

  On the base, loosen the 6 side screws that hold the Teflon cloth  

over the heating wire.

   

    3.

  Remove the Teflon cloth; may need to replace.

    

    4.

  The heating wire is held in place by two black plastic holders.   

Remove the screw on end with the timer knob.

    

    5.

 Remove the wire on that end.

    

    6.

 Pull the wire toward the timer knob end of the unit.

      

   7.

  Place an extra screwdriver or pen in the opening behind the other 

back holder; on the end with hinge. This will keep that piece in place.

    

    8.

 Remove the screw and the rest of the wire.

    

    9.

 Replace the wire.

    

 10.

 Screw the wire in place.

    

 11.

 Remove the extra screwdriver or pen.

 

12.  

Replace or put on a new Teflon tape.

    

 13.

 Tighten the 6 side screws.

    

 14.

 Plug in the unit.

    

 15.

 Check unit operation.

Содержание HS28271

Страница 1: ...______________________ 2 Safety Precautions_____________________ 2 Operating Instructions_________________ 3 Heating Adjustment Guide_____________ 3 Maintenance__________________________ 4 Parts Guide___________________________ 4 Instruction Manual ...

Страница 2: ...ta Sheet MSDS for any reagent you use with this sealer 4 Never operate the sealer in a hazardous or flammable environment 5 To protect against risk of electrical shock do not put sealer in water or other liquid 6 Blade is extremely sharp Keep hands hair clothing and other utensils away from blade during operation to reduce risk of injury to persons and or sealer 7 Always work in a manner which nei...

Страница 3: ...bar Press the handle down evenly 4 Heating is only applied while the light is on 5 When the light turns off either slide blade across to cut away excess material or lift handle to remove 6 Adjust the timer if needed INDICATOR POLYETHYLENE POLYPROPYLENE OR HIGH HEAT FILM 1 UNDER 0 06 mm 2 UNDER 0 1 mm 3 UNDER 0 14 mm 4 UNDER 0 2 mm UNDER 0 03 mm 5 UNDER 0 04 mm 6 UNDER 0 06 mm 7 UNDER 0 08 mm Heati...

Страница 4: ... need to replace 4 The heating wire is held in place by two black plastic holders Remove the screw on end with the timer knob 5 Remove the wire on that end 6 Pull the wire toward the timer knob end of the unit 7 Place an extra screwdriver or pen in the opening behind the other back holder on the end with hinge This will keep that piece in place 8 Remove the screw and the rest of the wire 9 Replace...