![Heathrow Scientific HS28271 Скачать руководство пользователя страница 4](http://html.mh-extra.com/html/heathrow-scientific/hs28271/hs28271_instruction-manual_2116890004.webp)
PART NUMBER
DESCRIPTION
PLUG TYPE
HS28270
220/230 VAC
U.K.
HS28271
220/230 VAC
Cont. Europe/Israel
HS28272
110/120 VAC
U.S./Taiwan/Japan
HS28280
Replacement Part Kit
—
2 extra blades, 2 extra heating elements and 2 extra PTFE cloths
Parts Guide
Maintenance
1.
Unplug the unit.
2.
On the base, loosen the 6 side screws that hold the Teflon cloth
over the heating wire.
3.
Remove the Teflon cloth; may need to replace.
4.
The heating wire is held in place by two black plastic holders.
Remove the screw on end with the timer knob.
5.
Remove the wire on that end.
6.
Pull the wire toward the timer knob end of the unit.
7.
Place an extra screwdriver or pen in the opening behind the other
back holder; on the end with hinge. This will keep that piece in place.
8.
Remove the screw and the rest of the wire.
9.
Replace the wire.
10.
Screw the wire in place.
11.
Remove the extra screwdriver or pen.
12.
Replace or put on a new Teflon tape.
13.
Tighten the 6 side screws.
14.
Plug in the unit.
15.