22
HeartSine recommends users perform
regular maintenance checks. The
recommended maintenance checks are:
Weekly
❏
Check the status indicator. If the
green status indicator is not flashing
every 5 to 10 seconds or if the red
status indicator is flashing or if you
hear beeping, a problem has been
detected. See ‘Troubleshooting’ on
page 25. The SAM 450P performs
a self-test routine at midnight GMT
every Sunday. During this self-test the
status light blinks red but returns to
green on successful completion of the
self-test routine. The self-test takes no
more than 10 seconds to complete. If
the status indicator continues to flash
red the SAM 450P has a fault (see
‘Troubleshooting’ on page 25).
Monthly
❏
If the device shows any signs of
physical damage, contact your
authorized distributor or HeartSine
Technologies directly.
❏
Check the expiration date of the SAM
450P Pad-Pak (see ‘Preparation’ on
page 14 for the location of the date).
If the date has expired, or is near
expiration, replace with a new Pad-
Pak or contact your local HeartSine
authorized distributor for a replacement.
If you hear a warning message when
you turn on your SAM 450P or if, for
any reason, you have suspicions that
your SAM 450P is not working correctly,
read the section ‘Troubleshooting’
on page 25.