Health O Meter PRO PLUS 2450KL Скачать руководство пользователя страница 19

 

  18

 

TROUBLESHOOTING 

Refer to the following instructions to check and correct any failure before contacting service 
personnel. 

SYMPTOM 

POSSIBLE CAUSE 

CORRECTIVE ACTION 

Scale does not turn on 

1.  Dead Battery 
2.  Faulty electrical outlet 
3.  Bad power supply 

1. Replace 

batteries 

2.  Use a different outlet 
3. Replace 

adapter 

Questionable weight or  
the scale does not zero 

1. External 

object 

interfering with the scale 

1.  Remove interfering object 

from the scale 

 

2.  The display did not show 

”0.0” before weighing 

2.  Ask the patient to get off 

the scale, zero the scale 
and begin weighing 
process again 

 

3.  Scale is not placed on a 

level floor 

3.  Place the scale on a level 

floor and begin weighing 
process again 

 

4.  Scale is out of calibration  4.  Check weight with known 

weight value 

 

5.  Improper Tare 

5.  Place the item to be tared 

on the scale ramps. Press 
REWEIGH. Once the 
weight of the item is 
displayed, press TARE. 
Place the patient and the 
tared item back on the 
scale. Press REWEIGH 
again. 

Weighing is performed but 
the display shows “weigh” 
and “reweigh” every few 
seconds; the weighing 
process takes too long and 
no weight is displayed. 

The patient is not sitting still 

Ask the patient to be still or 
you can change to live weight 
setting 

The display shows 
“Overload” message 

The load on the scale 
exceeds the capacity  
(600 Lbs) 

Remove the excess weight 
and use the scale according to 
its limits 

The display shows  
“LOW BAT” message 

The batteries are empty 

Replace batteries according to 
instructions 

The display shows  
“Load Cell Error” message 

There is a problem with one 
or more load cells or the load 
cell cable is disconnected. 

Check load cell cable 
connection at the display and 
platform assembly ports. If the 
problem is not corrected, refer 
to qualified service personnel 
to replace the defective load 
cell 

Содержание PRO PLUS 2450KL

Страница 1: ...TABLE WHEELCHAIR SCALE MODEL 2450KL BALANCE PORTATIVE PRO PLUSTM POUR FAUTEUIL ROULANT MODELE 2450KL BALANZA PORTABLE PRO PLUSTM PARA SILLA DE RUEDAS MODELO 2450KL OPERATION MANUAL MODE D EMPLOI MANUA...

Страница 2: ...l et le garder port e de main pour pouvoir le consulter tout moment Gracias por la adquisici n de este producto Lea por favor este manual cuidadosamente y mant ngalo a su alcance para una pronta refer...

Страница 3: ...d to provide accurate reliable and repeatable weight measurements and features that make the weighing process simple fast and convenient The scale is set up to use motion sensing weighing technology t...

Страница 4: ...y assembly on a flat stable table or on the floor 4 Open the scale and place it upside down on a level floor 5 Screw in the four plastic feet assemblies do not tighten Turn the scale assembly right si...

Страница 5: ...into the center channel in the display 12 Connect power cord to the power port in the display assembly and insert and secure the cable into the 2nd channel from the left in the display 13 Install the...

Страница 6: ...er NOTE If the set up procedure failed refer to the troubleshooting instructions If the problem is not corrected refer to qualified service personnel a To adjust the display backlight and or contrast...

Страница 7: ...r If necessary raise the scale by opening all four plastic feet 6 If the wheelchair weight is known enter the TARE value according to instructions in the Menu para 02 If the wheelchair weight is unkno...

Страница 8: ...ds Also used to scroll up in the menu MENU Enters the menu of the scale BMI Prompts entry of data to calculate the patient s Body Mass Index BMI PRINT Prints patient s data if printer is connected to...

Страница 9: ...scrolls to the bottom of the menu and presses the down button it will return to the top of the menu Figure 6 Main Menu 01 WEIGHT DISPLAY MODE only applies to pound values NOT the metric values The us...

Страница 10: ...ndicates the weight that has been tared out The number that is to be changed will flash and will move from the left to the right after the appropriate number was entered or by using the c key left and...

Страница 11: ...ory The maximum capacity of the scale is 270 files of different data 01 Auto Download Automatic Download is the default option and will transfer the value to the PC as soon as the patient steps off th...

Страница 12: ...stored in memory will be cleared 04 SCALE SETTINGS 01 Auto Hold Time The user can determine how long to display the weight reading once it is determined regardless of whether the patient remains on t...

Страница 13: ...adjust the beeping tone of the scale This tone should sound when the scale has determined the weight on the platform when a key is pressed after the scale is turned on at the end of self test or in th...

Страница 14: ...movement the scale will not lock on quickly to the weight as is the case in the motion sensing mode Press the REWEIGH button to operate the motion sensing mode and to determine the correct weight on t...

Страница 15: ...he batteries will have to be replaced NOTE In order to complete the battery test the scale must be powered by batteries only Unplug the scale from AC power source prior to battery test 02 USB Connecti...

Страница 16: ...llowing message will be displayed If after 10 seconds the requested command was not received the following message will be displayed If UI Error Failure is displayed refer to qualified service personn...

Страница 17: ...We recommend the use of this function to ensure that the patient s correct ID and height have been entered 03 Prompt For Height When this option is activated the user will be asked to enter the patien...

Страница 18: ...te correctly refer to qualified service personnel 1 Check overall appearance of the total scale for any obvious damage wear and tear 2 Inspect AC adapter for cord cracking or fraying or for broken or...

Страница 19: ...item to be tared on the scale ramps Press REWEIGH Once the weight of the item is displayed press TARE Place the patient and the tared item back on the scale Press REWEIGH again Weighing is performed b...

Страница 20: ...ion Please clear the scale 4 Please clear the scale platform and press ENTER Zero calibration Please wait 5 Please wait without touching or shaking the scale until the zero calibration process is fini...

Страница 21: ...20 EXPLODED VIEW OF SCALE continue...

Страница 22: ...NK PHIL HEAD SCREW M4 8 6 20 COUNTERSUNK PHIL HEAD SCREW M4 12 6 21 400117 PANDVIT CCS19 S8 C 1 22 LOCKING TV BAR 1 23 T BAR 1 24 DELRIN SPACER 1 25 CENTER SECTION A 1 26 2267501 0 PLASTIC FOOT NC3 8...

Страница 23: ...e or current use contrary to the operating instruction abuse including tampering damage in transit or unauthorized repair or alternations Further the warranty does not cover Acts of God such as fire f...