34
The different tables on the scale:
There will be two data tables on the scale. The first will store the sub ect record data (patient related
data that does not change) and the second will store the data recorded during the readings made.
Sub ect record data
Fields: ID, TARE, HEI HT
Readings
Fields: Inde (referring to the ID from the Sub ect Record Table), Weight, Date time
Sample and e planation of esc protocol
1. When the scale measures the weight of a patient and sends this over the
communications line to the PC, it will look like this:
SCALE PC
ESC R ESC I 1234
0 ESC W 0.
ESC H1
.
ESC T .
ESC B2
ESC Nm ESC E
2. When the PC requests the values of the stored weighings, it will look like this:
First the PC will ask the scale to send the data. We do this with the
et Sub ect Record Data and
Readings command.
SCALE PC
ESC
ESC E
Then the scale will reply with the stored data.
Reading (Command)
Patient ID = 1234567890
Weight = 90.98
Height = 178.70
TARE = 5.7
BMI = 28
Unit of measure = Metric
EOP
Get subject data
EOP
Direction of communication
Direction of communication