CLEANING

 

Proper care and cleaning is essential to ensure a long life of accurate and effective operation. Disconnect the scale from the battery charger. 
1.    Clean all external surfaces with a clean damp cloth or tissue. Mild soap and water solution may be used. Dry with a clean soft cloth. 
2.    Do not immerse the scale into cleaning or other liquid solution. 
3.    Do not use Isopropyl Alcohol or other solutions to clean the display surface. 
 

 

 

Refer to the following instructions to check and correct any failure before contacting service personnel. 

 

SYMPTOM

 

POSSIBLE CAUSE

 

CORRECTIVE ACTION

 

Scale does not turn on

 

1.  Dead battery 
2.  Faulty electrical outlet 
3.  Bad battery charger

 

1.  Attach battery charger to recharge battery. 
2.  Use a different outlet for the battery charger. 
3.  Replace battery charger.

 

Questionable weight or the scale 
does not zero

 

1.  External object interfering with the scale. 
2.  The display did not show 

“0.0” before 

weighing 

3.  Scale is not on a level floor 
4.  Scale is out of calibration

 

1.  Remove interfering object from the scale. 
2.  Ask the patient to get off the scale and begin weighing 

process again. 

3.  Place the scale on a level floor and begin weighing process 

again. 

4.  Check weight with known weight value.

 

The display shows 

“- 0F- ”

 

The load on the scale exceeds the maximum 
capacity (600 lbs / 270 kg)

 

Remove the excess weight and use the scale according to its 
limits.

 

 

 
 
 
 

This scale has been calibrated at the factory and does not require re-calibration prior to use.

 

The following procedure should be performed only to correct any error in calibration. For maximum accuracy, calibration should be done using 
certified weights. The recommended calibration weight is 600 lbs, however calibration can be performed using any weight no less than 44 lbs and no 
more than 600 lbs. 

 

1.    Turn the scale on. 
2.    At the same time press the TARE/ZERO and KG/LB buttons. 

“t” and the number representing the current automatic shut off value will 

appear on the display. 

3.    Press the TARE/ZERO button. 

“CAL” and then '600.0' will appear on the display. 

4.   To change the calibration weight, press the HOLD/RELEASE button to move the flashing indicator to the left. Then press the KG/LB button to set the     
       value to the desired calibration weight. 
5.    Press the TARE/ZERO button again. The display will show 'CAL 00'. 
6.    Press the TARE/ZERO button again. The display will show CAL-SP. Load 600 LB (or altered value) of calibration weight on to the scale and press the  
      TARE/ZERO button. The display will show "600.0" (or altered value). 
7.    Remove the weight from the scale. The scale returns to the normal weighing mode. 

 
 

 

ERROR NO.

   

DEFINITION

 

CHECK / REPLACE

 

Err 1

 

A / D value is lower when CAL

 

Scale is below minimum calibration weight. Calibration weight 
must be more than 44 lbs. The recommended calibration weight 

is 600 lbs.

 

--OF--

 

The load on the scale exceeds the capacity (600 lb / 270kg).

 

Remove excess weight and use the scale according to its 
limits.

 

Or-Err

 

High saturation state (High A / D)

 

The load cell cable is not connected properly. Check cables 
and mechanical connection. If problem persists, contact 
Health o meter Professional Customer Service 800-815-
6615.

 

Or-Err

 

Low saturation state (Low A / D)

 

The load cell cable is not connected properly. Check cables 
and mechanical connection. If problem persists, contact 
Health o meter Professional Customer Service 800-815-
6615.

 

UF

 

Battery is low.

   

Recharge battery. Overnight charge is required to fully charge 
battery.

 

TROUBLESHOOTING 

CALIBRATION PROCEDURE 

ERROR LIST 

Содержание 594KL

Страница 1: ...TING INSTRUCTIONS DIGITAL CHAIR SCALE MODEL 594KL Please register your product at www homscales com PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2016 P N UM594KL Rev 20160308 ...

Страница 2: ...eelchair to transport patients Do not bounce scale down stairs as doing so may damage internal parts Do not use the footrest as a step To maximize battery life do not leave charger plugged into the scale for extended periods GENERAL Health o meter Professional s Digital Chair Scale model 594KL uses sophisticated microprocessor technology This precision instrument is designed to provide accurate re...

Страница 3: ...or 3 seconds toggles between kilograms or pounds 4 TARE ZERO Zeros the scale prior to weighing or sets a tare value 5 ON OFF Turns scale ON and OFF The 594KL is equipped with a power switch located on the side of the display When the switch is moved to the On I position it is supplying power to the scale even when the display head is off To charge the battery this switch must be in the On I positi...

Страница 4: ...ghts The recommended calibration weight is 600 lbs however calibration can be performed using any weight no less than 44 lbs and no more than 600 lbs 1 Turn the scale on 2 At the same time press the TARE ZERO and KG LB buttons t and the number representing the current automatic shut off value will appear on the display 3 Press the TARE ZERO button CAL and then 600 0 will appear on the display 4 To...

Страница 5: ...ct use on improper voltage or current use contrary to the operating instruction abuse including tampering damage in transit or unauthorized repair or alternations Further the warranty does not cover natural disasters such as fire floods hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province...