Health O Meter 553KG Скачать руководство пользователя страница 7

 

OPERATING INSTRUCTIONS 

IMPORTANT: To prevent falls or injury, place the scale securely far from the edge of the 
counter, table or other flat surface used to hold the scale. Do not let the scale touch the 
wall or any other object near or around the scale while it is in use. 
 
Weighing a Patient 

1.  Make sure there is nothing on the tray. 
2.  Press the 

ON/OFF

 

button to turn the display assembly on. “88:8:8.8” followed by “0.0” will 

appear on the display.  

3.  Place the baby on the scale so that its weight is evenly distributed on the tray. The scale will 

begin to display weight information and will lock on the patient’s weight when the stable 
weight of the patient is determined.  Depending on the movement of the patient on the scale, 
it may take several seconds for the scale to lock onto the weight.  

4.  When a stable weight is calculated, a small ring 

along with the word “HOLD” will appear on 

the display and the weight of the baby will lock.  

4.  Note: 

If the display shows “OVER” you have exceeded the capacity of the scale.         

(553KL:44 lb / 20 kg; 553KG: 20 kg)  

Caution: To prevent patient injury, the patient must be attended throughout the entire 
weighing event.   

5.  Remove the baby from the scale. T

he scale will continue to display the baby’s weight.  

6.  To return the scale to 0.0, press the 

RELEASE

 button and the scale will be ready to weigh 

again.  

Note:

 

If “Lo” appears on the display, it indicates that the battery power is low. It does NOT 

indicate a problem with the weight measurement. 

 
Zero/Clear (TARE) Function 

The tare function allows you to place other items on the scale while weighing the baby, without 
affecting the baby’s weight reading.  
1.  Place the item you want to tare on the weighing tray. 
2.  Press and release the 

ZERO/CLEAR

 button

. The word “TARE” will appear on the display 

and the display will return to 0.0.  

3.  Place the baby on the scale with the tared item and weigh normally. The scale will only 

display the baby’s weight.  

Note:

 If there is no weight on the scale and it does not read 0.0, press the 

ZERO/CLEAR

 button 

once to return the display to zero.  
 

Changing the Weight Measuring Unit (553KL Only) 

1.  Press the 

ON

 button to power the scale on. 

2. 

Wait for “0.0” LB or “0.0” KG to appear on the display. 

3.  Press the 

kg/LB

 button to toggle to the desired weight measuring unit. 

 

 
Locking the Weight Measuring Unit in KG (553KL Only) 

Note: 

The weight measuring unit can be locked to only measure in KG or LB. 

1.  Press the 

ON

 button to power the scale on. 

2.  Wait for 

“0.0” LB or “0.0” KG to appear on the display. 

3.  Press and hold the 

kg/LB

 

button for approximately 20 seconds until “KG” flashes in the 

lower right corner of the display. This locks the scale in KG mode and will not display weight 
in pounds. 

4.  To release the KG lock, press the 

kg/LB 

button for approximately 20 seconds until “LB”, 

“OZ” and “KG” flash on the display. The user can now toggle the weight measuring unit 
between LB and KG. 

Содержание 553KG

Страница 1: ...Scale 44 lb 0 5 oz 20 kg x 0 01 kg User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM553KL_553KG Rev20180111_EngOnly 232791 IEC 60601...

Страница 2: ...Connectivity Disposal 5 Powering the Scale Set Up 6 Operating Instructions 7 Maintenance 8 Calibration 9 Troubleshooting 10 Warranty 11 Note This scale has been factory calibrated and does not requir...

Страница 3: ...Use of an unspecified adapter will void the warranty and can pose a serious safety hazard Prior to using this scale inspect the AC adapter cord for cracking fraying or for broken bent plug prongs Pri...

Страница 4: ...l Operating temperatures 50 F to 104 F 10 C to 40 C Storage Specifications 32 F to 122 F 0 C to 50 C Atmospheric Pressure 500hPa 1060hPa Humidity 10 90 RH Physical Dimensions Tray Size Length 13 330 m...

Страница 5: ...nnectivity Solutions This connectivity helps minimize transcription errors when recording patient data For further information regarding how to connect this scale to other electronic devices please co...

Страница 6: ...adapter is being used the battery icon will not be displayed Measuring Tape For your convenience a measuring tape is included with the scale You may affix the tape to either the bottom or side of the...

Страница 7: ...e If Lo appears on the display it indicates that the battery power is low It does NOT indicate a problem with the weight measurement Zero Clear TARE Function The tare function allows you to place othe...

Страница 8: ...be consumed Items such as clothing or towels place or dropped on the scale may cause it to activate To prevent run down of the batteries do not store or leave anything on the scale Cleaning Proper ca...

Страница 9: ...can calibrate CAL followed by XXXXX Press the ZERO CLEAR button C 5 will flash on the display and then return to 0 C 5 then 0 Place 5 kg on the scale and when the calibration number stabilizes press t...

Страница 10: ...Remove the patient from the weighing tray zero the scale and begin weighing process again 3 Scale is not placed on a level surface 3 Place the scale on a level surface and begin weighing process again...

Страница 11: ...nadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get War...