background image

3

Paper Chart Installation

Use the following procedure to install or 
replace a chart in the recorder.
1. Press the raised ends located on either 

side of the case lid to remove the lid.

2. Replace the pen cap on pen from the 

pen cap holder below the recording 
chart. Remove the recorded chart if 
present.

3. Place a new chart on the chart hub, 

making sure that the center strip is 
intact.

4. Set the time by inserting a coin into 

the chart hub (pen screw) and turning 
clockwise to the correct time (and day 
for 7-day units) relative to the time 
index clip located at the middle right 
hand side of the recorder.

5. Remove pen cap from pen and put it 

on pen cap holder below the recording 
chart. 

6. Replace the lid. At the end of one cycle 

(24 hours or 7 days, depending on your 
recorder), simply repeat steps 1-6, 
replacing the used chart with a new 
one.

Pen Installation

Disposable fiber-tipped pens are installed 
in this recorder. The pens have a self-
contained ink supply and are housed in 
a replaceable cartridge. Normally, a new 
cartridge pen will give many months of 
recording, depending on usage. When 
replacement becomes necessary, use the 
following procedure.
1. Carefully lift the pen arm.
2. While holding the pen arm, use your 

thumb and forefinger to grasp the pen 
cartridge by the finger grips near the tip 
of the pen.

3. Gently slide pen cartridge straight off 

pen arm.

4. With the pen tip facing the chart, thread 

a new pen cartridge onto the pen arm. 
Slide the pen cartridge back until you 
feel the notch in the pen arm seat 
securely around the plastic cartridge of 
the pen tip.

5. Lower the pen on the chart.
6. If the pen arm is bent in the process, 

carefully bend the arm back so the pen 
contacts the paper.

7. Remove the pen cap and store on the 

pen cap holder on the lower part of the 
unit. 

Quick Start 

1.  These recorders come with 1 AA 

battery installed.

2.  Set switch on back of unit to the 

appropriate motorspeed: 7 day or 24 
hour.

3.  Replace chart paper, with a new chart.
4.  Set the time by inserting a coin into 

the chart hub and turning clockwise 
until the correct hour is referenced by 
the time indicator. (Time indicator is 
directly to the right and below the pen 
tip.)

5.  Remove protective pen cap.
6.  The chart movement is self-starting 

and runs continuously. Allow unit 
to stabilize in environment for 15 
minutes.

3

© Health Care Logistics, Inc. 2010

Call Free: 1.800.848.1633 or 1.888 HCL INTL

® 

(425 4685)

 

• 

Fax Free: 1.800.447.2923

Website: www.HealthCareLogistics.com • Email: [email protected] 

Отзывы: