● Seal up cleaning, disinfection and sterilization waste to protect the environment.
7.4 Daily Maintenance of Products
● Check power frequently, and DO NOT use the device with impropriety power.
● Use soft, dry mull to clean device. DO NOT use dripping cloth to wipedevice.
● There should be at least 1 hour working time of the power supply and functional check time
every month. DO NOT deposit the device without power more than 6 months.
● Contact the professional service people if fail to repair and give a brief introduction of
equipment fault state.
Chapter 8 Simple fault check and elimination
8.1 Check
● Check whether the computer power is normal and check whether the probe USB plug is
plugged into the computer USB port and has a good contact.
8.2 Trouble Shooting
● Trouble Shooting
(
shown in the following table
)
number
Fault
Methods of Trouble Shooting
1
Open system software and prompt
device configuration fails.
1. Check computer output-voltage
;
2. Check the probe into the USB port
;
2
The
screen
is
disturbed
by
intermittent line and snow.
1. Check whether there is high power equipment
running around;
2. Check whether the system is disturbed by
electromagnetic field around;
3. Check that the plug of the probe is in good
contact with the USB port.
3
Image on the screen is not clear.
1. Adjust GAIN
;
2. Adjust Brightness and Contrast
;
3. The near field image is not clear. Adjust the
top 3 of 9 sliders which are below the GAIN
4
.
The far field image is not clear. Adjust the