38
<00001> 11:00 01/02/2015
25.2℃ 390μS/cm 15.0μS/cm
18.2MΩ.cm 12.4L 00012.4L
<00002> 12:00 02/02/2015
25.0℃ 400μS/cm 15.2μS/cm
18.2MΩ.cm 15.6L 00028.0L
Data Description:
The first line from left to right: product water dispenses times, time and date.
Second line from left to right: water temperature, feed water conductivity and pure water
conductivity.
The third line from left to right: resistivity of ultra-pure water, volume of dispensed and
total volume of dispensed.
Press "Disp. /
▲
" or "Flush /
▼
" keypad to scroll the screen to check other time of the
water data record.
Press "Menu" keypad to return to the previous menu, or press "Menu" keypad
continuously to exit the maintenance setting, the system restart.
Cartridges replacement records
Operate in accordance with this section to enter data record menu, press "Disp. /
▲
" or
"Flush /
▼
" keypad to select replacement records. White on black means that it has been
selected. As shown below.
Water
Record
Replacement
Record
Press "Stop /
" button to confirm, LCD displays each record for water producing. As
shown below.
10:20 30/06/2015
PF
02
11:00 30/06/2015
RO
01
11:20 30/06/2015
DI
04
11:50 30/06/2015
MF
01
Data Description:
From left to right: replacement time, cartridge type, cumulative number of replacement
times.
Press "Disp. /
▲
" or "Flush /
▼
" keypad to scroll to see other time for the consumables