Application
WTX120
A4500-1.1
HBM: public
77
Hysteresis
In Below level mode, the hysteresis specifies the amount by which the level
must exceed the threshold for the limit status - and so also the digital output -
to be reset.
8.2.3
Mode: Outside band
Weight
Time
Out
Lower band
Band span
Lower band / Band span
In Outside band mode, Lower band and Band span define a band. Outside that
band, the relevant limit status is set. The limit is then likewise set on a precon
figured digital output (e.g.
output 3 GW1
).
As soon as the weight is back in the band, the limit status is reset.