Calibration
WTX110
A4774-1.0
HBM: public
183
Function/Secondary selection
Key
Menu
Maximum Y 15.000
Enter the maximum angle in the Y direction for
which the correction will still be performed. If the
value is exceeded the weight indicator is hidden.
Incline mode Correct
Operating principle of the optical tilt sensor:
Correction
The weight is corrected according
to the inclination.
Check
:
A check determines whether the
angle is within the valid range. The
display is hidden outside of this
range. There is no correction
(function like the tilt switch).
11.9.2
Incline calibration
Menu
Key
Function/Secondary selection
Incl.
calibration
0-Adj.Incl.Sensor?
Linearization of the optical tilt sensor with
unloaded scale
N
No
Y
Yes
X= 0.754 Y= -0.110?
Display of the current X, Y angle
Measure ...
Measurement running. The mean values of
multiple measurements are determined.
Offset X0
0.706
Display/enter the correction value in X direction
Содержание WTX110
Страница 1: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 3: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 10: ...8 A4774 1 0 HBM public WTX110 17 Geo values 212...
Страница 55: ...Installation WTX110 A4774 1 0 HBM public 53 Fig 4 12 Positioning of ALI option on the WTX110 motherboard...
Страница 217: ...Operating Manual Bedienungsanleitung English Deutsch WTX110...
Страница 224: ...8 A4774 1 0 HBM public WTX110 17 Geowerte 217...
Страница 259: ...Installation WTX110 A4774 1 0 HBM public 43 Abb 4 6 Positionierung DIO Modul SIO auf der WTX110 Hauptplatine...
Страница 271: ...Installation WTX110 A4774 1 0 HBM public 55 Abb 4 12 Positionierung Alibi Speicher auf WTX110 Hauptplatine...