Electrical connection
16
A02335_03_Y00_00
HBM: public
VK20A
5.1.4
Electrical connection of the voltage supply (St7)
Strip terminal/
terminal
Function
St7
/1
O
/2
control signal (external trigger 0 V; )
I
/3
control signal (external trigger 5 V … 30 V)
I
/4
O
/5
Supply voltage VK20 24 V (14 … 30 V)
I
/6
Supply voltage 0 V;
I
Tab. 5.3
Connector pin assignment St7
1)
O=Output; I=Input
2)
Linked internally with St7/Pin6. Can be bridged with St7/Pin2 if necessary.
3)
Linked internally with St7/Pin5. Can be used for triggering the control signal if necessary.
4)
In conjunction with MGC
plus
through St3/1.