Mechanical installation
5.1
Important measures for installation
•
Treat the transducer gently.
•
Make sure that there is a rigid base when measuring compressive forces.
•
The force-introduction surfaces must be totally clean and fully bearing.
•
Comply with the engagement depths for threaded rods or knuckle eyes.
•
Do not overload the transducer.
•
Either during installation or immediately afterwards, the transducer should
be bridged by a 50mm
2
stranded copper wire (highly flexible earthing cable
EEK from the HBM line). The cable is attached by screws both above and
below the transducer. This prevents welding current flowing over the
transducer and welding the force introduction point.
WARNING
If there is a risk of breakage through overload on the transducer
and thus a risk to persons, additional safety measures are to be
taken.