93
Connection
HBM
QuantumX
6.5.1
MX410 pin assignment
So that insertion or removal of a transducer connection can be unmistakably identified, Pin 4
and Pin 9 in the connector plug must be bridged! If this bridge is missing, no measurement
values will be recorded at the connection!
1
4
5
6
9
11
15
Bridge
Fig.6.7:
Pin arrangement of connection plug, view from the solder side
Pin
Connection
1
TEDS (+)
2
Bridge Excitation voltagen (
−
)
3
Bridge Excitation voltage (+)
4
Always connect with Pin 9! (Plug identification)
5
Measurement signal (+)
6
TEDS (
−
)
7
Sense lead (
−
)
8
Sense lead (+)
9
Signal ground
10
Measurement signal (
−
)
11
Active sensor supply (
−
)
12
Active sensor supply (+)
13
Current input
"
30 mA (+)
14
Voltage input 10 V, IEPE (+)
15
Reset external charge amplifier
The analog output can be tapped via BNC. Please see chapter 7 ”Functions and outputs” for
information about the configuration.
NOTE
Many HBM transducers are fitted with 15-pin D-SUB connectors (2 rows). The adapter
cable 1-KAB416 can be used for connection to the 3 row D-SUB-15HD device
connectors of the MX840. Pins 4 and 9 are already bridged in this adapter cable.
Содержание QuantumX MX1609
Страница 1: ...I3031 8 0 en Operating Manual ...
Страница 2: ...d ...
Страница 31: ...31 Housings HBM QuantumX Fig 4 2 Amplifier MX1609 P in IP65 housing ...
Страница 34: ...34 Housings HBM QuantumX a f 2 5 Fig 4 7 Mounting the CASEPROT ...
Страница 36: ...36 Housings HBM QuantumX Fig 4 10 Attaching the covers ...
Страница 38: ...38 Housings HBM QuantumX Fig 4 13 Close the lever Fig 4 14 Connected housings ...
Страница 99: ...99 Connection HBM QuantumX Connection of the thermocouple plug in the miniature design ...
Страница 144: ......
Страница 180: ...179 Support HBM QuantumX ...