Electrical connections PMX
PMX
A4354-2.2
HBM: public
53
8
Electrical connections PMX
8.1
Plug technology and clamping areas
All PMX plug-in cards (PX401, PX455, PX460, PX878)
are delivered as standard with easy to fit plug terminals
using Push-In technology. Plug terminals can also be
delivered with a screw-on system as an option.
Push-In technology
The clamping area is 0.2mm2 (AWG24) to 1.5mm2
(AWG16). If several conductors are to be connected to a
terminal, the line cross-section must be adapted
accordingly. End sleeves (without plastic collars, 10mm)
should be used on the strands to connect the wires to the
terminals.
The shield of the transducer cable must be connected to
the ground connection provided on the PMX multipoint
connector, in accordance with HBM Greenline informa
tion
.
Important
The grounding terminal on the PMX is not a protective
ground (optional connection).
The measurement system is fitted with an automatic
current limiter for each device card, as well as for the
PMX basic device.
Содержание PX401
Страница 1: ...Operating Manual English PMX...
Страница 104: ...Electrical connections PMX 104 A4354 2 2 HBM public PMX...
Страница 136: ...Startup 136 A4354 2 2 HBM public PMX This takes you to the device overview Now you can measure make settings and observe...
Страница 141: ...Startup PMX A4354 2 2 HBM public 141 Network settings...
Страница 238: ...Internal calculation channels 238 A4354 2 2 HBM public PMX Appendix Details of the glyphs in example A...
Страница 239: ...Internal calculation channels PMX A4354 2 2 HBM public 239...
Страница 241: ...Internal calculation channels PMX A4354 2 2 HBM public 241...
Страница 246: ...Internal calculation channels 246 A4354 2 2 HBM public PMX...
Страница 264: ...Internal calculation channels 264 A4354 2 2 HBM public PMX...
Страница 272: ...Parameter sets measurement programs 272 A4354 2 2 HBM public PMX...
Страница 295: ...Communication with a control system PMX A4354 2 2 HBM public 295...
Страница 297: ...Communication with a control system PMX A4354 2 2 HBM public 297...
Страница 300: ...Communication with a control system 300 A4354 2 2 HBM public PMX...
Страница 376: ...CODESYS V3 Soft PLC WGX001 only 376 A4354 2 2 HBM public PMX...
Страница 378: ...CODESYS V3 Soft PLC WGX001 only 378 A4354 2 2 HBM public PMX...
Страница 381: ...CODESYS V3 Soft PLC WGX001 only PMX A4354 2 2 HBM public 381...
Страница 392: ...CODESYS V3 Soft PLC WGX001 only 392 A4354 2 2 HBM public PMX Dialog Updating the PMX device...
Страница 489: ...Object directory PMX A4354 2 2 HBM public 489...
Страница 523: ...Diagnosis and maintenance PMX A4354 2 2 HBM public 523 Network settings...