![HBM MP60DP Скачать руководство пользователя страница 17](http://html.mh-extra.com/html/hbm/mp60dp/mp60dp_operating-manual_2111904017.webp)
17
PME
−
MP60DP
HBM
A0623-6.1 en/de/fr
Status 1
Bits
Name
Meaning
0
Ovfl
measured values overflow
1
AOutOvfl
analogue output overflow
2
ScalErr
scaling defective
3
EEPROMErr
EEPROM (parameter set) defective
4
Limit1
status of limit switch 1
5
Limit2
status of limit switch 2
6
Limit3
status of limit switch 3
7
Limit4
status of limit switch 4
8
PAR1
active parameter set bit 1
9
PAR2
active parameter set bit 1
10..14
res
reserved
15
MViO
Measured value in order
1)
(if bit 0,2,3=0)
Tab 6.5:
Contents of status 1
1)
Meaning of MViO:
NOR operation of: MesswOvfl, SkalErr, EEPROMErr.
MesswOvfl is the OR operation of ADCOvfl, HardwOvfl, GrossOvfl, NetOvfl
The parameter set number is coded in 2 bit binary:
Bit 8 Bit 9
Parameter set no.
0
0
1
1
0
2
0
1
3
1
1
4
Status 2
Double status word 2 returns detailed error flagging.
Bits
Name
Meaning
0
HardwOvfl
hardware overflow
1
ADCOvfl
ADC overflow
2
GrossOvfl
gross signal overflow
3
NetOvfl
net signal overflow
4
AOutOvfl
analogue output overflow
5
MaxOvfl
maximum overflow
6
MinOvfl
minimum overflow
7
NegOvfl
overflow in negative direction
8
Limit1
status of limit switch 1
9
Limit2
status of limit switch 2
10
Limit3
status of limit switch 3
11
Limit4
status of limit switch 4
12
ScalInError
scaling input invalid
13
ScalOutError
scaling output invalid
14
GainError
nominal value exceeded
15
InitError
works calibration defective
16
TransducerError
transducer error
17..31
res
reserved
Tab 6.6:
Contents of status 2