Manual parameterization and configuration
46
A00760_09_E00_00
HBM: public
ML77B
11
Manual parameterization and configuration
11.1 Parameterisation
The parameterisation telegram has the following entries:
Parameterisation entry
no.
Meaning
Contents
0
Device parameterisation
1
Channel 1
2
Channel 2
3
Channel 3
4
Channel 4
...
...
...
16
Channel 16
The data format that has been set applies for all measured values exchanged
in the cyclic data traffic. The definition of decimal places is only relevant for the
Signed16 and Signed32 formats (Example: With 3 decimal places defined, 2.0
mm will be transmitted as integer value 2000). The data format selected also
affects the length of the input data (Signed16 = 1 word per analogue value,
Signed32 and Float = 2 words per analogue value).
The targeted enabling of the required control bits in the control word enables all
functions that are not required to be protected against an accidental activation
in the event of an error; otherwise, for example, the zero point that has once
been adjusted could be lost. To enable, set bit to 1.