10.3.4
Dual
-
window trigger mode
The dual
-
window trigger mode is a more sophisticated version of the window
trigger mode. Now both levels are used as an arm/trigger/disarm level. This
allows the trigger detector to react to a dip in both directions.
A
A
A/T/D
2
1
A/T/D
A/T/D
A/T/D
Figure 10.7:
Dual
-
window trigger mode
A
Trigger
Diagram
1
shows one situation, whereas Diagram
2
shows the other situation
with the same settings. In this mode, the following conditions determine the
trigger result:
l
Level crossing in opposite set direction = arm level
l
Level crossing in set direction = disarm when other level is armed
l
Level crossing in set direction = trigger when level is armed
Since this is true for both levels, a “dip” in both directions is detected, as shown
in Diagrams
1
and
2
Relevant settings for this mode:
l
Mode: dual
-
window
l
Primary level: any value within the input range
l
Secondary level: any value within the input range
l
Direction: positive or negative for primary level; secondary level is
automatically set to the opposite direction
l
Hysteresis: any relevant value is used for both levels.
GEN3i
I3763-3.1 en HBM: public
151
Содержание GEN3i
Страница 1: ...Portable Data Recorder GEN3i English User Manual I3763 3 1 en HBM public...
Страница 4: ...4 I3763 3 1 en HBM public...
Страница 51: ...33 46 7 70 IEC 61010 1 2010 IEC EN 50110 1 EN 50110 2 50 120 GEN3i I3763 3 1 en HBM public 51...
Страница 119: ...Figure 8 2 Acquistion control Detail A Record B Stop C Pause D Trigger GEN3i I3763 3 1 en HBM public 119...
Страница 822: ...Once the computer name has been entered correctly click Next Figure E 14 Computer name GEN3i 822 I3763 3 1 en HBM public...
Страница 1158: ...1158 I3763 3 1 en HBM public...
Страница 1159: ......