![HBM FIT 7A Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/hbm/fit-7a/fit-7a_operating-manual_2111833011.webp)
Mechanical installation
FIT
®
7A
A03924_02_YI0_00
HBM: public
9
Mounting
Thread
Min. property class
Tightening torque
On mounting plate
M8
10.9
25N
⋅
m
For load application
M6
10.9
14N
⋅
m
1)
Recommended value for the specified property class. Please comply with the screw
manufacturer's instructions with regard to screw dimensions.
Mounting plate
(not included among
the items supplied)
Connection for valve coupling plug
(included among the items supplied)
Socket 1
Socket 2
*)
*)
Version E only
Mounting plate flatness <0.02 mm
Load application
center of platform
Platform
(not included among
the items supplied)
Fig. 4.1
Load application and installation
Important
The flatness of the mounting plate surface to which the load cell is attached
must be better than 0.02mm.
4.3
Connection for pressure compensation
The load cell housing is fitted with a connection for pressure compensation.
Important
Changes in air pressure will also affect the measurement result, if you do not
connect the valve coupling plug (the load cell is then sealed).