![HBM BU18 Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/hbm/bu18/bu18_mounting-instructions_2111861009.webp)
9
BU18
HBM
A2306
−
1.3 en/de
3
Mounting
3.1 Removing from the carrying case
The system is delivered fully assembled. To remove it from the carrying case,
remove the upper attachment from the carrying case. The complete unit can
then be lifted at the central ring nut or alternatively by the three eyebolts at the
side.
ATTENTION
Before lifting, you must check that the ring nut is screwed right down
onto the centering bolt and that this is screwed right down onto the
lifting anchor. The eyebolts at the side must also be screwed right in
(use the wrench).
Use only approved hoisting gear with sufficient carrying force.
3.2 Mechanical assembly
The BU18 comprises a mounting base, the transducer, transducer mounting
base, pendulum and thrust piece assemblies and the top plate with the guide
element. The guide element is not designed to be removed.
The BU18 mounting base and the C18 mounting bases, as well as the
transducers themselves, are connected by centering bolts. These bolts are
interchangeable. A central lifting anchor with centering bolts is attached in the
top plate and three side eyebolts are attached to the BU18 mounting base for
transporting the BU18.