AED
CONNECTING INTERFACES AND I/OS
7.4
AED9301B
7.4.1
Connection to the PROFIBUS
To next device
S2
S3
Bus termination
Bus
disconnection
on
off
on
off
A
A B
B
B
A
KL3
A
B
From the master
S5
S4
Address settings
Fig. 7.12
Pin assignment and switch positions for PROFIBUS
The PROFIBUS DP is standardized as per IEC61158 / EN50170. The AED is a PROFIBUS
DP slave as per DIN19245-3. The PROFIBUS is electrically isolated from the measuring
system and the supply voltage. The maximum possible bit rate is 12Mbit/s. Set the node
address using S4 and S5:
Address = S4 + 10 • S5 (permitted address range: 3...99)
Set the address in the deactivated state; it is read in by the AED when the supply voltage
is applied. The factory-set address is 03.
PROFIBUS connection KL3 has four terminals, so that you can route it to the next bus
node. Use shielded, twisted-pair cables as connecting cables.
Activate the bus termination at both ends of the PROFIBUS cable. Bus termination may
only be activated at the end points of the bus (max. 2 termination resistors active). If nec
essary, slide switch S2 (bus termination) on the AED to on. The factory setting is off.
You can disconnect the module from PROFIBUS for the purpose of diagnosis: Switch S3
to off. The factory setting is on.
Содержание AED 9101D
Страница 1: ...AED ENGLISH DEUTSCH Operating Manual Bedienungsanleitung ...
Страница 3: ...AED Transducer Electronics 9101D 9201B 9301B 9401A and 9501A ENGLISH DEUTSCH Operating Manual ...
Страница 53: ...AED Aufnehmer Elektroniken 9101D 9201B 9301B 9401A und 9501A ENGLISH DEUTSCH Bedienungsanleitung ...
Страница 103: ......
Страница 104: ...HBK Hottinger Brüel Kjaer www hbkworld com info hbkworld com A05803 01 X00 00 ...