Assembly Instructions M8103-EN 

 

M8103-EN   2020-01 

 

2/4

 

 

  

 

with product group Thermo-5 

 

1.

 

Proceed as follows in order to cool down the unit and empty 
the mould (mould evacuation): 

  Display menu page 

Functions

  Select the function 

Cooling

 and activate with the 

 key.

 

  If the function 

Mould evacuation 

is available, select it and 

activate it with the 

 key. 

  The activated function is indicated with the 

 symbol. 

 

  Units HB-100X_4 and HB-100/140/160Z_4 without function 

Mould evacuation

: Loosen the drain connection (G) slowly until 

the unit is depressurized. 

 

with product group Treat-5 / Clean-5 

 

2.

 

Switch on Emptying tank function as follows: 

  Display menu page 

Functions

.

  Select the parameter 

Emptying tank

 and activate with the 

 

key. 

The active function is indicated with the 

symbol. 

 

3.

 

Main switch off, remove the plug from the mains and empty 
the unit. 

 

 

4.

 

Remove the cover from the unit (see Instruction Manual 
Chapter Maintenance). 

  

5.

 

Loosen the screws in the front panel and hinge it down. 

 

6.

 

To remove pressure sensor (BP 1, 2), proceed as follows: 

  Disconnect the pressure sensor cable to GIF-51 (A 4), remove 

the cable ties and completely remove the cable. 

 

Pressure sensor system (BP 1) 

 connector X 31.1 

  Pressure sensor main line (BP 2) 

 connector X 31.2 

  Remove the hydraulic connection to the pressure sensor. 

Remove the pressure sensor from the mounting bracket using a 
hexagon nut. 

 

 

7.

 

To mount the pressure sensor (BP 1, 2), proceed as follows: 

  Screw the pressure sensor with hexagon nut to the  

mounting bracket and install the hydraulic connection. Lay the 
pressure sensor cable through the cable duct and connect to 
GIF-51 (A 4). Lay the pressure sensor cable through the cable 
duct and connect to GIF-51 (A 4). 

 

Pressure sensor system (BP 1) 

 connector X 31.1 

 

Pressure sensor main line (BP 2) 

 connector X 31.2 

 

 

 

  

 