20
2
Operation
DC voltage measurement / V=
Set the appropriate range using the selector switch. Connect the black test lead to the "
COM
"
jack and the red test lead to the "
V/Ω
" jack. Connect test leads to the test object. Read off
measurement result from the display. The voltage polarity is also displayed.
Measurement range
Resolution
Input resistance
Overvoltage protection
Accuracy
600 mV
0.1 mV
>1000 M
Ω
250 Vrms
±0.7 % of the
measurement
range + 2 digit
6 V
1 mV
>10 M
Ω
750 Vrms
60 V
10 mV
600 V
100 mV
AC voltage measurement / V~
Set the appropriate range using the selector switch. Actuate the "
SEL
button" and select AC.
Connect the black test lead to the "
COM
" jack and the red test lead to the "
V/Ω
" jack. Connect
test leads to the test object. Read off measurement result from the display.
Measurement range
Resolution
Input resistance
Overvoltage protection
Accuracy
600 mV
0.1 mV
>1000 M
Ω
250 Vrms
±0.8 % of the
measurement
range + 3 digit
6 V
1 mV
>10 M
Ω
750 Vrms
60 V
10 mV
600 V
100 mV
Frequency range: 10 Hz - 1.2 kHz
^
hazet.com •
]
Electrical energy may cause serious injuries!
Always wear personal protective equipment!