rev.
09.09.2009
cjr
©2009
HAUSMANN
INDUSTRIES,
INC.
HAUSMANN INDUSTRIES, INC.
Model 1440 - Special Wrench Instructions
ATTACHING AND DETACHING ACTUATOR PLUGS TO CONTROL BOX
The 1/4” plugs must be inserted into the control unit until the O-rings are
no longer visible, and turned approx 30° to the right as far as the stop by
means of special plastic wrench to lock. If this is not done, strain relief and
water tightness are not guaranteed.
DETACHING POWER PLUG AND FOOT SWITCH PLUG FROM CONTROL BOX
The two lines may only be detached from the control box by means of special plastic wrench with which the
brackets can be separated, by pulling the plug (see illustrations). The brackets can break if they are opened too
wide. Strain relief and watertightness are no longer guaranteed. Note: Some force is still required to remove plug
from Control Box.
ATTACHING POWER PLUG AND FOOT SWITCH
PLUG TO CONTROL BOX
When attaching Power Plug and Foot Switch Plugs
to Control Box ensure that the plugs are inserted
in the correct position to prevent damage to the
control unit and /or the plug. (Note plug shape.)
Power Plug Port
Foot Switch Port
Actuator Port