HAND SWITCH
Operation of the system is by means of an LD-011 Hand Control Switch (mounted on black
wooden support post near one end of platform).
PATIENT WEIGHT CAPACITY
400 lbs.
NOTE: Handrail motors are designed to electrically adjust in height , and to be stable under use
when at their fixed height level.
Motors are not designed to "lift" full weight of patient from
sitting to standing position.
Not designed for continuous use.
Intermittent 10%; 30 seconds ON
270 seconds OFF
Thermal overload: Unit will shut down automatically if motor overheats due to continuous use.
Unit will restart automatically after "cool off" peiod of up to 15 minutes.
OPERATING METHOD
2
↥
2004 Hausmann Industries, Inc.
11/03/04
Push Force (1 Lift Column)
Stroke
Speed
110V, 60Hz, 2.5 Amps
All electrical components are U.L. recognized components, furniture grade.
There is sufficient wheelchair clearance for most 16" and 18" wide seat wheelchairs, but
insufficient clearance for 20" or wider seat wheelchairs.
TECHNICAL DATA
1000 N
15" (29" to 44" high)
2.1"/second