III
Do not use or operate this product at sites where there is a risk of explosion
.
The electrical connection of this product and the connection of the inlet and outlet
pipes should be carried out by professionals.
This product should not be in direct contact with the patient. Place this product outside
the range of at least 1.5 meters (radius) in the treatment area.
If an external printer is configured for this product, the printer should not be placed in
a location where there is a risk of contact with liquid.
Emergency door
opening device
Do not use this device to open the door under normal conditions.
Sewage pretreatment
and sterilization
Pretreatment and sterilization of instruments and fabrics should follow the regulations
of the instrument and fabric manufacturers
.
Strictly follow the relevant standards for disinfection and sterilization
,e.g.WS506‐
2016,WS310‐2016.
Use packaging materials or containers that have been confirmed by their manufacturer
for sterilization.
Sterilization termination
This product terminates abnormally at certain stages of the program,high temperature
steam will emerge from the chamber after opening the door.
The loaded item may not be sterilized when the sterilization is abnormally terminated.
Please note the status of the program showed on the display when it terminates.If
necessary, the loaded items should be repackaged and sterilized.
Remove the loaded item
Do not use strong door opening.
Use the tray handle to take out the tray
.
Do not touch the sterilized items with bare
hand,
including
chamber and door cover.
Check if the loaded items is packaged completely after sterilization.If the package is
damaged
,p
lease pack instrument or fabric again
.
Delivery and transportation
At least 2 people need to move or deliver this product
.
Use proper belts
.
The shipping box provided by the manufacturer of this product must be used for
transportation.
Sterilization load
The load must be sterilized with allowed steam by the manufacturer before it can be
sterilized.
maintenance
Maintenance of this product must be performed by a professional authorized by the
manufacturer.
Please maintain the product at the time intervals specified in this manual
.
Must use the accessories supplied by the manufacturer of this product
.
Содержание Q50B
Страница 1: ...Small Steam Sterilizer User Manual Internal Water Tank Q50B Q70B 2460 H Tmed Wuxi Co Ltd V01 version...
Страница 9: ...2 Fig 1 1 front view 1 2 2 Side view Fig 1 2 side view Door handle Power switch Display...
Страница 39: ...32 Piping system schematic Appendix 3 Electrical schematic diagram...
Страница 40: ...Small steam sterilizer electrical schematic diagram...