MODEL HPM-2002-OBE
QUICK START GUIDE
HASTINGS INSTRUMENTS
READ ME FIRST!
1. Unpacking and Inspection
Unpack and inspect all items for any obvious signs of damage due to
shipment. Immediately advise the carrier who delivered the shipment if
any damage is suspected.
Transducer
OBE Module
D-connector
4. Gauge Installation
Transducer tube may be installed in any orientation. However, if
condensation is likely to occur, then port should be orientated
downward.
When installing 1/8” NPT style transducer tube, use the 7/16” wrench
flats as shown.
Attach cable.
With the vacuum chamber at atmosphere, turn on the power supply
(typ24 VDC. Gauge is now reading pressure. The pressure
in Torr as read by channel 1 is given by, P = 100 x V where V is the
voltage between pins 5 and 6.
For best accuracy, the gauge should now be zeroed. Pump the
vacuum system down to low (10
-6
Torr) pressure if possible. Ideally
the gauge should be operated in this condition for one hour.
Place the HPM-2002-OBE in the “CAL” mode by using the “SELECT”
button. Using a small flat head screwdriver, rotate the “ADJUST”
rotary encoder until the unit flashes between “0.0” and “-0.0”.
Return to the “RUN” mode by using the “SELECT” button.
*Similary the pressure in mTorr as read by channel 2 is given by
P = 100 x V where V is the voltage between pins 7 and 6.
7/16” wrench
flats