background image

 
At the conclusion of the sterilization cycle, a buzzer will sound and the "Sterile" Light will 
illuminate 

(see figure 1 below)

. Accordingly, it is not necessary to observe the pressure indicator 

gauge during the sterilization cycle.  
 

COMPLETING THE CYCLE

 

BE CERTAIN CONDENSATE TRAY IS FULLY INSERTED IN PLACE TO PREVENT 
CONDENSATE SPILLAGE. When the Timer Buzzer and Sterility Indicator Light (see figure 1 
below) signal completion of the sterilization cycle, turn the control knob to "Depressurize" (see 
figure 1 below).  
 
An efficient system of coils assures condensation of virtually all chemical vapor in the chamber 
in approximately 25 seconds. AS A SAFETY FEATURE, THE DOOR IS DESIGNED TO BE 
OPENED ONLY AFTER THE CHAMBER IS EXHAUSTED. 
 
The contents may be immediately removed. NO DRYING IS REQUIRED.  

 

HELPFUL SUGGESTIONS

 

Your Chemiclave is unsurpassed as an instrument for sterilization in the professional office, 
clinic or hospital. Treat it accordingly. Following are useful comments:  

 

PARTS

 

The 

Door Gasket

 should be replaced when damage, deterioriation, or wear may prevent proper 

seal. To maximize the life of the the gasket, leave door closed but unlatched when the 
Chemiclave is not in use.  

 

MAINTENANCE

 

Exterior metal surfaces of your Chemiclave should be cleaned daily with Harvey Metal Polish. 
Care should be exercised to prevent polish from entering valve and timer stem openings in the 
solution reservoir. 
 
Because of the detergent action of Vapo-Steril, debris from improperly cleaned instruments may 
accumulate in the chamber and instrument tray. Harvey Metal Cleaner should be used 
periodically as needed to remove such debris. AT LEAST WEEKLY CLEANING IS 
RECOMMENDED. 
 

 

Отзывы:

Похожие инструкции для CHEMICLAVE 5000