At the conclusion of the sterilization cycle, a buzzer will sound and the "Sterile" Light will
illuminate
(see figure 1 below)
. Accordingly, it is not necessary to observe the pressure indicator
gauge during the sterilization cycle.
COMPLETING THE CYCLE
BE CERTAIN CONDENSATE TRAY IS FULLY INSERTED IN PLACE TO PREVENT
CONDENSATE SPILLAGE. When the Timer Buzzer and Sterility Indicator Light (see figure 1
below) signal completion of the sterilization cycle, turn the control knob to "Depressurize" (see
figure 1 below).
An efficient system of coils assures condensation of virtually all chemical vapor in the chamber
in approximately 25 seconds. AS A SAFETY FEATURE, THE DOOR IS DESIGNED TO BE
OPENED ONLY AFTER THE CHAMBER IS EXHAUSTED.
The contents may be immediately removed. NO DRYING IS REQUIRED.
HELPFUL SUGGESTIONS
Your Chemiclave is unsurpassed as an instrument for sterilization in the professional office,
clinic or hospital. Treat it accordingly. Following are useful comments:
PARTS
The
Door Gasket
should be replaced when damage, deterioriation, or wear may prevent proper
seal. To maximize the life of the the gasket, leave door closed but unlatched when the
Chemiclave is not in use.
MAINTENANCE
Exterior metal surfaces of your Chemiclave should be cleaned daily with Harvey Metal Polish.
Care should be exercised to prevent polish from entering valve and timer stem openings in the
solution reservoir.
Because of the detergent action of Vapo-Steril, debris from improperly cleaned instruments may
accumulate in the chamber and instrument tray. Harvey Metal Cleaner should be used
periodically as needed to remove such debris. AT LEAST WEEKLY CLEANING IS
RECOMMENDED.