T +44 (0)1709 377172 F +44 (0)1709 377173
20
ROUTINE
MAINTENANCE
SEAT CUSHION
The seat cushion cover, which is made from waterproof and vapour permeable material,
should be kept clean. Take care to avoid puncturing cover with sharp objects whilst
performing the maintenance checks:
1
Remove cover and inspect for damage, tears or staining, which could lead to
contamination of the internal parts.
2
Check that the zips are sound and in good working order.
3
Check that all connectors are fitted properly to prevent leaking of air.
4
Check that all cells are attached to the base sheet by the pop fittings provided.
5
Check the stitching on the straps and the seams to ensure no tearing or fraying
has occurred.
These checks should be carried out at each decontamination process, i.e.
between patients or patient occupancy and weekly for longer term patients.
PUMP
1
Check the pump casing for cracks or other damage that could be dangerous.
2
Check the power cord (ensure there are no bare wires).
3
Check for the correct function of the low pressure light, disconnect the pipes from
the seat cushion into transport mode, the pump should indicate low pressure
within a few seconds
If any faults are detected report to your distributor for replacements to facilitate repairs.
COMPONENTS
Check air cells and seat cushion interior for signs of damage or contamination, e.g.
staining or fluid ingress at each decontamination process, i.e. between patients or
patient occupancy (or weekly for longer term patients).
The individual cells can be wiped clean with a mild antiseptic solution.
All cells are replaceable and can be sourced from Harvest Healthcare.
POWER UNIT
Disconnect the power unit from the electricity supply before carrying out maintenance, repairs
or cleaning.
Check all electrical connections and power lead for signs of wear and damage.
The power unit can be wiped down with detergent, disinfectant solution or wipe*. Do not use
solvents. Unsuitable for sterilisation.
Содержание BLENHEIM ACTIVE SEAT CUSHION
Страница 2: ...T 44 0 1709 377172 F 44 0 1709 377173 2 ...
Страница 13: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 13 OVERVIEW Internal view of alternating cells ...
Страница 29: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 29 NOTES ...
Страница 30: ...T 44 0 1709 377172 F 44 0 1709 377173 30 NOTES ...
Страница 31: ...E sales harvesthealthcare co uk www harvesthealthcare co uk 31 NOTES ...