OPERATION
Audible / Visual Warning conditions are indicated by a flashing red display
accompanied by an audible warning. In each case the user should respond by turning
the Power Unit’s switch off and investigating the cause.
1. High Pressure Warning
If the system pressure exceeds 100mmHg this warning will be activated. This may be
triggered by, for example, a kinked Umbilical Hose or another person suddenly sitting
on the mattress.
2. Low Pressure Warning
If the system pressure falls below 15mmHg this warning will be activated. This may be
triggered by, for example, incorrect fitting of the air inlet connector, opening of the CPR
Valve or a leak in the Mattress due to a cut or puncture.
3. Mains Failure Warning
If mains power is lost then all LED lights will turn off. In such an event, the Mains
Failure audible warning will sound.
AUDIBLE/ VISUAL WARNING