17
H
a
r
v
a
r
d
A
p
p
a
r
a
t
u
s
P
u
m
p
'1
1
'
P
l
u
s
S
y
r
i
n
g
e
P
u
m
p
Publication 5411-001-REV-E
Queries:
DIA
Returns diameter value units in mm
RAT
Returns rate value set in current range units
VOL
Returns current accumulated infused volume, units in ml or µl
depending on range
VER
Returns model and version number of firmware
TAR
Returns target volume, units in ml or µl depending on range
value format: nnnn.nnn
The returned value is an 8 character string with leading zeros converted to SP
characters (ACSII 32 decimal). The fifth character is a decimal point (ASCII
46 decimal)
Queries with string response:
RNG
Returns range message (character string either: ML/H ML/M
UL/H UL/M)
Error responses:
CR LF ? CR LF prompt
Unrecognized command
CR LF OOR CR LF prompt
Entered value in out of range
Pump Chain
Pump Port 1 (in)
-
Computer control side
Pump Port 2 (out) -
Connection for remainder of pump chain
Baud Rate
-
300, 1200, 2400 or 9600
Word Size
-
8
Parity
-
none
Stop Bits
-
2
Appendix D: Serial Communication