Single Syringe Pump

MA1 70-2208

Dual Syringe Pump

MA1 70-2209

Single Syringe Pump with Serial Communication

MA1 70-2211

Dual Syringe Pump with Serial Communication

MA1 70-2212

Model ‘11’ Plus

Syringe Pump

User’s Manual

Publication 5411-001-REV-E

Содержание MA1 70-2208

Страница 1: ...A1 70 2208 Dual Syringe Pump MA1 70 2209 Single Syringe Pump with Serial Communication MA1 70 2211 Dual Syringe Pump with Serial Communication MA1 70 2212 Model 11 Plus Syringe Pump User s Manual Publication 5411 001 REV E ...

Страница 2: ...ng will be offered for our products which are within the scope of the WEEE Directive Products falling under the scope of the WEEE Directive available for sale after August 13 2005 will be identified with a wheelie bin symbol Two Categories of products covered by the WEEE Directive are currently exempt from the RoHS Directive Category 8 medical devices with the exception of implanted or infected pr...

Страница 3: ...ted Turn Pump ON 8 Function Keys and Run Indicator 8 Entering Syringe Size 9 Entering Flow Rate Range 9 Entering Flow Rate 9 Press Run 9 Check Syringe 9 Advanced Features Volume Mode 10 Power Failure 10 Display Intensity 11 Changing Rates 11 Infuse Withdraw Switch 11 Maintenance 11 Protecting Small Fragile Syringes 12 Remote Communication 12 Appendices A Specifications 13 B Table of Popular Syring...

Страница 4: ...BILITY OR FITNESS FOR A PARTICULAR USE OR OF ANY OTHER NATURE Some states do not allow this limitation on an implied warranty so the above limitation may not apply to you If a defect arises within the two year warranty period promptly contact your local distributor or Harvard Apparatus 84 October Hill Road Holliston Massachusetts 01746 1388 using our toll free number 1 800 272 2775 valid only in t...

Страница 5: ... integrity of the device use only one of the following power supplies Ault Inc Model PW118 Output 12 Vdc 1 5 A Input 100 250 Vac 50 60 Hz 5A Cui Inc Model SA06N12 V Ouptut 12 Vdc 2 0 A Input 100 240 Vac 50 60 Hz 8A Globtek Inc Model GT 4201D 12 Output 12 Vdc 1 66 A Input 100 240 Vac 50 60 Hz 0 6A Use Proper Line Cord Use only the line cord shipped with the product and make sure line cord is certif...

Страница 6: ...rate the product Contact qualified service personnel to perform inspection Observe all Warning Labels on Product Read all labels on product to ensure proper usage Enviromental Conditions Indoor use only Temperature 4 C to 40 C 40 F to 104 F Humidity 20 to 80 RH Well Ventilated Room Altitude up to 2000 m Mains Voltage Fluctuation not to Exceed 10 of Nominal Transient Overvoltage Category II Pump is...

Страница 7: ...on data The Power Failure Mode can be set to either turn the pump off in the event of power failure or to resume pumping when power resumes Features Bright Display and Easy To Use Interface A two line 16 character vacuum fluorescent display along with six membrane keys make this a most attractive but powerful easy to use syringe pump Only two entries required to start pumping syringe Inside Diamet...

Страница 8: ...t and Rear Views Pusher Block 4 Keypad Bronze Button 3 LED Light Shaft Collar 8 Syringe Block 5 Syringe Retainer 1 Dual RS 232 Connections Option Pusher Block bracket 2 Syringe block bracket 6 Power Input jack On Off Switch Guide Rods Infuse Limit switch adjuster 7 Infuse Withdraw Switch Standard only with RS232 option ...

Страница 9: ...Release the Syringe pusher block 4 by pressing the bronze button 3 on the side of the pusher 2 While holding the bronze button in slide the pusher to the right 3 Raise the spring loaded syringe retainer 1 and swing it out of the way 4 Lay the loaded syringe in the V shaped syringe block 5 5 Swing the syringe retainer 1 so it holds the syringe in place 6 Move the pusher so it makes contact with the...

Страница 10: ...rline cursor below the digit or character on the display moves one step to the right It is used in conjunction with the ascending and descend ing keys When it the display shows the desired the correct value the set key will advance right to the next digit L LM M The ascending and descending keys are used to change the numbers on the display L L Up key makes numbers increase M M Down Key makes numb...

Страница 11: ...your flow rate units while in the SET UNITS mode by pressing the L L or M M keys to scroll the four flow rate choices ml min µl min ml hr µl hr Once the desired flow rate units are displayed press the ENTER key to return to the main SET CONFIG mode Press SET or ENTER again to put your desired flow rate units into memory and return to the RATE VOL CONFIG menu These units will be the same for infuse...

Страница 12: ...nse you can restart the pump at the place you stopped by pressing the RUN key again To exit the volume dispense mode set the target volume to 00 00 or turn off and on the pump via the main power switch NOTE In the event of a power failure the actual dispensed volume and the target volume are not retained in memory This means that while in volume dispense mode if a power failure occurred the pump w...

Страница 13: ...e the pump is running As soon as the ENTER key is pressed the pump will change to the new flow rate To change rates from the keypad while in volume mode the pump must be stopped first 5 Infuse Withdraw Switch Operation standard with RS232 option For customer convenience a two position rocker switch is included in the Pump 11 Plus Advanced for changing from infuse to withdraw on the fly When the pu...

Страница 14: ...ptions Use the L L or M M key to scroll the CONFIG options until you reach the SET ADDRESS Press the SET key again moving the underline cursor to the right Use the L L or M M key to scroll the two digit number fields Use the L L or M M key and the SET key to display a unique pump address from 00 to 99 Press the ENTER key to select the desired pump address Exit the CONFIG mode by pressing the ENTER...

Страница 15: ...n led Nonvolatile Memory Stores diameter rate and configuration settings Maximum Force 8 kg 17 lb Maximum Pressure 8 p s i with 50 60 cc syringe 100 p s i with 1 ml syringe Drive Motor 0 9 step angle motor Step Rate Minimum Maximum 1 pulse in 27 6 sec 200 steps sec Pusher Advance Step 0 33 µm pusher advance per motor step Pusher Travel Rate Minimum Maximum 0 7 µm min 15 mm min Dynamic Range 1 to 1...

Страница 16: ... 300 500 3 260 1000 4 610 Terumo Size Diameter 3 cc 8 95 mm 5 13 00 10 15 80 20 20 15 30 23 10 60 29 10 Sherwood Monoject Plastic Size Diameter 1 cc 4 65 mm 3 8 94 6 12 70 12 15 90 20 20 40 35 23 80 60 26 60 140 38 40 Cadence Science Inc Formly Popper Sons Inc Perfektum Glass Size Diameter 0 25 cc 3 45 mm 0 5 3 45 1 4 50 2 8 92 3 8 99 5 11 70 10 14 70 20 19 58 30 22 70 50 29 00 100 35 70 SGE Scien...

Страница 17: ... 0 0003 4 765 0 0001 0 0794 250 µl 2 30 0 7218 9 999 0 0121 197 0 0 0008 11 82 0 0001 0 1970 1 000 µl 3 26 1 451 9 999 0 0242 395 7 0 0015 23 75 0 0001 0 3959 1 ml 4 61 2 900 9 999 0 0484 791 8 0 0029 47 50 0 0001 0 7918 2 5 ml 7 28 7 232 9 999 0 1206 1 974 0 0073 118 4 0 0002 1 974 3 ml 8 66 10 24 9 999 0 1706 2 794 0 0103 167 6 0 0002 2 794 5 ml 10 30 14 50 9 999 0 2413 3 952 0 0145 237 1 0 0003...

Страница 18: ...rt infuse forward direction STP Stop motor CLV Clears volume accumulator to zero CLT Clears target volume to zero dispense disabled REV Start Reverse Direction Commands with numbers MMD number Set syringe diameter units are mm Rate is set to zero after MMD command ULM number Set flow rate and range units are microliters per minute MLM number Set flow rate and range units are milliliters per minute...

Страница 19: ... value format nnnn nnn The returned value is an 8 character string with leading zeros converted to SP characters ACSII 32 decimal The fifth character is a decimal point ASCII 46 decimal Queries with string response RNG Returns range message character string either ML H ML M UL H UL M Error responses CR LF CR LF prompt Unrecognized command CR LF OOR CR LF prompt Entered value in out of range Pump C...

Страница 20: ... Publication 5411 001 REV E Appendix E Dual RS 232 Daisy Chain Connector Cables Part No Description 70 2022 RS232 serial cable 6 ft long 9 pin d sub on one end and RJ11 on the other end 72 2478 Daisy Chain Cable 6 ft 55 7760 Daisy Chain Cable 2 ft 72 8340 USB to Serial Adapter ...

Страница 21: ...p p a r a t u s P u m p 11 P l u s S y r i n g e P u m p Publication 5411 001 REV E Appendix F PC Connection 3 RJ11 RJ12 6 RJ12 6 x 6 View at Pump 5 4 6 View of Adapter View of Cable Front View Model 11 Plus Syringe Pump ...

Страница 22: ...pparatus Inc Manufacturer s Address 84 October Hill Road Holliston Massachusetts 01746 U S A Type of Equipment Syringe Pump Model No Model 11 Plus I the undersigned hereby declare that the equipment specified above conforms to the above Directive s and Standard s Place United States of America Date September 06 2002 Declaration of Conformity Beth Bauman VP Engineering Operations Signature Full Nam...