Recording Waveforms Using the 

Harvard Modular Universal Oscillograph

Numbers in brackets, e.g. (1), (2) etc. refer to Fig. 1 on page 3.

Letters in brackets, e.g. [A], [B] etc. refer to Fig. 5 on page 9 and 

Fig. 12 in the Modular Universal Oscillograph Instruction Manual.

1)

Have you read the complete Operating Procedure section?

2)

Check that the Isolated Preamplifier is fitted with batteries

(see page 12).

3)

Refer to Fig. 2, 3 or 4 on pages 5, 6 and 7 to apply the elec-

trodes to your subject.

4)

Connect the electrodes to the Isolated Preamplifier using

either the 3-lead input cable or the ECG limb cable.

5)

With reference to the Harvard Modular Universal Oscillograph

Instruction manual section 3.4.0, follow steps 1 to 13 inclu-

sive.

6)

Plug the Isolated Preamplifier to the Isolator Interface.

7)

Switch the Isolator Interface on and check that the battery

indicator (1) on the Preamplifier is in the green zone.  If it is

not, refer to page 12.

8)

Set the Preamplifier’s AC/DC switch (5) to the DC position

and set the filter switch to 15 kHz.

9)

Set the Preamplifier's GAIN switch (6) to 10 and select posi-

tion I on switch (2).

10)

Set the Oscillograph's Direct-OFF-Interface switch [AC] to

INTERFACE, the AC/DC switch [AA] to DC, the chart speed

control [T] to 2.5, and start chart drive by pushing chart drive

lever [M] to ON.

11)

Adjust the Oscillograph’s GAIN control [W] to about 1.  The

pen will move on the chart.

12)

Adjust the Preamplifier’s DC LEVEL control (3) until the pen

returns to its central position on the chart.  At this point you

should see a small waveform on the chart.

8

Содержание 50-9513

Страница 1: ...84 October Hill Road Massachusetts 01746 Tel 508 893 8999 Fax 508 429 5732 e mail bioscience harvardapparatus com Web Site harvardapparatus com Operating and Maintenance Instructions Harvard Isolated...

Страница 2: ...illograph 8 9 Use with the Harvard Student Oscillograph 10 11 Problems 12 Service Information 12 13 Customer Contact Addresses 13 Diagrams Figure 1 Isolated Preamplifier Controls 3 Figure 2 ENG Electr...

Страница 3: ...ing transmitted Introduction The Harvard Isolated Preamplifier has been designed for ECG EKG EMG and ENG EOG recording with subject isolation The front end of the preamplifier is a battery operated op...

Страница 4: ...V Filtering 3 position filter switch 30 Hz 3 dB down 150 Hz 3 dB down 15 kHz 3 dB down Gain Selectable x10 x100 x1000 Other Facilities DC level control battery level indicator choice of DC or AC coupl...

Страница 5: ...needle electrode ALWAYS STERILIZE THE ELECTRODES BEFORE AND AFTER USE ON EACH SUBJECT to reduce the chance of skin infections or blood carried diseases being transmitted Always ensure that the electro...

Страница 6: ...ing of the reference electrode or zero volt electrode This electrode within reason can be placed anywhere 5 For EMG just place the electrodes over the group of muscles to be monitored see Fig 3 below...

Страница 7: ...e 0 volt reference Essentially an ECG can be recorded from any part of the body but unless the conventional positions are used the interpretation of the results would be difficult For more guidance pl...

Страница 8: ...7 AVL I AVR II III AVF V1 V2 V4 V5 V6 V3...

Страница 9: ...ction 3 4 0 follow steps 1 to 13 inclu sive 6 Plug the Isolated Preamplifier to the Isolator Interface 7 Switch the Isolator Interface on and check that the battery indicator 1 on the Preamplifier is...

Страница 10: ...ier s FREQ CUT control 4 If you are performing a 5 lead ECG you may select the various limb configurations using switch 2 Calibrate the waveform using the Oscillograph s the CAL 0 1 switch Z This will...

Страница 11: ...raph Instruction manual section 3 4 0 follow steps 1 to 13 inclu sive 6 Connect the Isolated Preamplifier to the INTERFACE socket S on the Student Oscillograph 7 Check that the battery indicator 1 on...

Страница 12: ...control 4 If you are performing a 5 lead ECG you may select the various limb configurations using switch 2 Calibrate the waveform using the Oscillograph s the CAL 0 1 switch Q This will cause the pen...

Страница 13: ...on the Isolated Preamplifier 4 If the Isolated Preamplifier is sited close to a main power lead then move one or the other or preferably both as far apart as possible Service Information Fitting or R...

Страница 14: ...address below stating the nature of the fault and the approximate date of purchase if possible ENGLAND Harvard Apparatus Ltd Telephone 0732 864 001 Fircroft Way Fax 0732 863 356 Edenbridge Telex 9529...

Отзывы: