25
Publication 5420-009 REV1.0
Appendix B: Using Pressurized
and/or Anesthetic Gases
The VentElite was designed to operate utilizing gas sources of atmospheric pressure. When
pressurized, constant flow devices are used with the VentElite (such as regulators, flowmeters,
and house gas supplies) measures must be taken to ensure no damage is done to the unit.
We recommend the use of our connection kit when using a high pressure gas source with the
VentElite, or any of our ventilators. The connection kit prevents the buildup of pressure from
constant flow devices such as those referenced above. Below is a schematic of the connection
kit. The appropriately sized connection kit for the VentElite is the 73-4872 connection kit.
Pin (DB15) Signal
Description
1
HOLD_OUT
Output for Hold event
2
SIGH_OUT
Output for Sigh event
3
SYNC_OUT
Output showing Run state
4
I/E_OUT
Output showing Inhale/Exhale state
5
SIGH_I
Input to cause Sigh event
6
HOLD_I
Input to cause Hold event
7
RUN/STOP
Input to Start/Stop ventilator
8
IN1
Trigger Input (inactive)
9-15
GND
GND
Appendix C: Digital I/O
Connection Kit
Compressed Gas Source
• Vaporized
• Compressed Gas Tank
• House Gas Source
Evacuation
• Filter Canister
• Active Scavenging
• Open to Room Air
Arrows
Indicate
Gas Flow
Inspiratory and Expiratory
Pathway
Subject