12. Technical data
Measuring method:
Oscillometric and Korotkoff’s principle of
blood pressure measurement
Display range:
0 – 300 mm Hg
Measuring range:
Systole (SYS): 50 – 250 mm Hg
Diastole (DIA):
40 – 180 mm Hg
Pulse:
40 – 160 beats / minute
Technical measuring precision: Cuff pressure: +/- 3 mm Hg
Pulse: +/- 5% of displayed pulse rate
Power supply:
4 x 1.5 V Mignon alkaline-manganese
(AA/LR06) batteries or optional
HARTMANN Tensoval mains adapter
Battery capacity:
Tensoval duo control:
> 1,300 measurements
Tensoval duo control large:
> 1,000 measurements
Inflation pressure:
approx. 190 mm Hg
Automatic switch-off function: 3 minutes after end of measurement
Cuff:
Standard cuff 22 – 32 cm
Large cuff 32 – 42 cm
Pressure release valve:
Electronically controlled linear valve
Memory capacity:
2 x 30 measurements and mean value
Operating conditions:
Ambient temperature: +10 °C to +40 °C
(+ 50 °F to + 104 °F)
Relative humidity:
15 – 90%
Storage / transport conditions: Ambient temperature: –20 °C to +50 °C
(– 4 °F to +122 °F)
Serial number:
See battery compartment
13. Power supply, disposal note
and safety information
13.1 Batteries, mains adapters
and disposal
m
The four high-quality batteries
included with the device guaran-
tee about 1,300 measurements.
Use only high-quality batteries
(see specification in Chapter 12
“Technical data”). If you are
using less efficient batteries we
can no longer guarantee 1,300
measurements.
m
Never mix old and new batteries
or batteries made by different
manufacturers.
m
Immediately remove exhausted
batteries.
m
If you do not intend to use the
device for a longer period of
time, you should remove the
batteries to prevent possible
leakage.
m
Please help protect the environ-
ment: Do not dispose of batte-
ries in household waste. Use
designated collection points or
municipal collection/recycling
centres to dispose of hazardous
household waste.
m
Information concerning the
disposal of electronic devices
(private households):
This symbol on products
and/or accompanying docu
ments means that spent elec-
tronic products must not be mixed
with regular household waste. Take
these products to designated col-
lection points where they are
accepted free of charge for proper
treatment, reclamation and recy-
cling. Proper disposal is in the
interests of environmental protec-
tion and prevents possible harmful
effects on people and the environ-
ment that may result from inappro-
priate handling of wastes. More
detailed information about your
nearest collection point is available
from your local council office.