8. Explanation of error messages
Error
Possible causes
Remedy
Monitor cannot be
switched on
Batteries are missing,
incorrectly inserted or low/
empty.
Check batteries and insert
four identical new batteries if
necessary.
Power supply unit not correctly
connected or faulty.
Make sure the power supply is
connected to the socket on the
back of the device.
Cuff will not inflate Cuff connector plug is
incorrectly positioned in the
socket of the monitor.
Check connection between cuff
plug and connection socket.
Wrong cuff type connected.
Check that only the approved
Veroval
®
duo control cuff and
the corresponding connector
were used.
The measuring signals could
not be detected correctly or
at all.
Too few pulse beats or pulse
too weak.
Unusual measurement time or
result.
General measurement error.
Check whether the cuff has
been applied correctly. Do
not talk or move during the
measuring procedure.
Cuff does not inflate or does
not inflate fast enough.
The cuff was applied too firmly
or too loosely.
Unusual function of the
pressure sensor.
Apply the cuff so that two
fingers fit between the cuff and
the upper arm.
Air tube is not correctly
connected to the monitor.
Check that the connector plug
is correctly positioned. If this
error occurs frequently, you
should use a new cuff.
IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918.indd 29
17.10.18 18:04
Содержание 9255020
Страница 1: ...IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 1 17 10 18 18 03 ...
Страница 2: ...2 English EN IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 2 17 10 18 18 03 ...
Страница 40: ...40 Türkçe TR IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 40 17 10 18 18 04 ...
Страница 79: ...79 Notlar TR IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 79 17 10 18 18 04 ...
Страница 81: ...81 AE ﻋﺮﺑـﻲ IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 81 17 10 18 18 04 ...
Страница 118: ...118 ﻣﻼﺣﻈﺎت AE IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 118 17 10 18 18 04 ...
Страница 120: ...120 繁體中文 HK IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 120 17 10 18 18 04 ...
Страница 158: ...158 IFU_Veroval_dc_SK2_0307661_EN_TR_AE_CN_200918 indd 158 17 10 18 18 04 ...