![HART Proline Promag 55 Скачать руководство пользователя страница 129](http://html1.mh-extra.com/html/hart/proline-promag-55/proline-promag-55_manual_601405129.webp)
Device Functions Proline Promag 55
10 Block SPECIAL FUNCTION
Hauser
129
10.1
Group ADVANCED DIAGNOSTICS
Introduction
The optional software package "Advanced Diagnostics" (F-CHIP) can be used to detect changes to
the measuring system at an early stage, e.g. as a result of build-up (coating), abrasion and corrosion
at the measuring electrodes. Such factors cause a reduction in accuracy in normal cases or lead to
system errors in extreme cases.
With the aid of diagnostic functions it is possible to record the following diagnostic parameters
during operation:
• Decay times of test pulses at the measuring electrodes
• Electrode potentials at both measuring electrodes
• Volume flow value (before applying the test pulses)
By analysing general trends of these diagnostic parameters, deviations of the measuring system from
a "reference condition" can be detected at an early stage, allowing for countermeasures to be taken.
Measurement of the decay time constant of test pulses (Fig. 1):
Monitoring both measuring electrodes makes it possible to detect the formation of build-up at an
early stage. To do this, a defined voltage pulse (U
B
) with a pulse width (t
p
, typically 1 to 20 ms) is
applied periodically at an electrode and its decay time constant (
τ
R
) is measured. The decay time
constant is a function of the condition of the measuring electrode in question.
a0006131
Fig. 1:
Schematic curve of the decay time constant of a voltage pulse at a measuring electrode.
U
0
= zero voltage, U
B
= voltage of the test pulse for the coating detection, t
p
= pulse duration,
τ
R
= decay time constant,
t
E
= recovery time
Measurement of electrode potentials:
The measuring electrode potential is influenced by various factors, for example by solids, air
bubbles, inhomogeneities in the fluid, sudden pH changes, mechanical damage or corrosive
changes. Therefore, monitoring the electrode potentials provides information about the specified
disturbance factors.
Measurement of the volume flow (immediately before applying the test pulses):
What is meant here by "volume flow" is the volume flow value that is acquired immediately before
the test pulses are applied to the measuring electrodes. This value serves as another basis for the
interpretation of decay time constants or electrode potentials with regard to coating formation,
abrasion or corrosion.
U [V]
U
B
t
p
t
R
t
E
t
U
0
Содержание Proline Promag 55
Страница 2: ...Device Functions Proline Promag 55 2 Endress Hauser ...
Страница 5: ...Device Functions Proline Promag 55 Contents Endress Hauser 5 ...
Страница 164: ...14 Index Device Functions Proline Promag 55 164 Endress Hauser ...
Страница 166: ...www endress com worldwide BA120D 06 en 10 09 71104957 FM SGML6 0 ...