STATS:
Various statistical values are displayed numerically for the given sen-
sor. Statistical values to display can be selected using the FIELD DATA menu
function (Section 7.2.3).
T/H
: Temperature and humidity are displayed numerically for the given sensor.
T GRAPH:
Temperature history is displayed on a graph for the given sensor.
The temperature (vertical) scale, time (horizontal) scale, and number of grid di-
visions can be set for the zone using the GRAPH SCALE menu function (Sec-
tion 7.2.4).
H GRAPH
: Relative humidity history is displayed on a graph for the given
sensor. The relative humidity (vertical) scale, time (horizontal) scale, and num-
ber of grid divisions can be set for the zone using the GRAPH SCALE menu
function (Section 7.2.4).
Graphs include time tick marks that occur at hourly intervals for time scales 24
hours or less or at daily intervals for time scales greater than 1 day. Time tick
marks are labeled at intervals with the hour or day, labels being placed below
and to the right of tick marks. Tick marks with labels are drawn with short con-
tinuous lines, and tick marks without labels are drawn with dotted lines.
Graphs are replotted when a different display layout is selected, a display lay-
out's settings are changed, or graph scale settings are changed. Data for the
graphs are extracted from recorded data in memory. If recording is disabled,
graphs cannot be replotted and will appear empty until new data is collected.
7.2.3
Field Data
The FIELD DATA function allows you to select a type of statistic for any of the
fields in a STAT type zone. This may be password-protected (see Section 7.5.4,
Password). Select the display layout on which to adjust the statistics field type
and then select the zone to adjust. This function can only be used with display
layouts that contain STATS zones. There are eight fields per zone. Any field
can be set to the following field types: DATE, TIME, ID, SERN, TEMP, T
RATE, T AVE, T STD, T MIN, T MAX, T SPR, T N, T RMAX, T ALAR, T
DELT, RHUM, H RATE, H AVE, H STD, H MIN, H MAX, H SPR, H N, H
RMAX, H ALAR, H DELT, DEWP, HEAT, REC CH, BATT, MEAS, or
BLANK. See Table 2 for a definition of each of these parameters.
Statistics fields show running statistics of the measurements taken since the last
statistics reset. The statistics are reset automatically after power up, when the
STATS RESET function is used (see Section 7.3.2.4), or when the time reaches
1620 the “DewK”
37
7 Menu Functions