Chapter 5 - Block Transfers
5-7
Block Read
Command Number
4: Read Tare Value
TABLE 5-7: BLOCK READ COMMAND NUMBER 4: READ TARE VALUE
Block Read
Command Number
5: Calibration
Parameters
BLOCK READ COMMAND NUMBER 4: Read Tare Value
START
WORD DEFINITIONS:
#WORDS
WORD
Command number: A value of 4 (decimal)
bit 0
0
bit 1
0
bit 2
1
bit 3
0
bit 4
0
bit 5
0
bit 6
0
bit 7
0
bits 8 - 15
Reserved for future use
1
0
Tare Value
2
1
TOTAL NUMBER OF WORDS
3
BLOCK READ COMMAND NUMBER 5:Calibration Parameters
START
WORD DEFINITIONS:
#WORDS
WORD
Command number: A value of 5 (decimal)
bit 0
1
bit 1
0
bit 2
1
bit 3
0
bit 4
0
bit 5
0
bit 6
0
bit 7
0
bits 8 - 15
Reserved for future use
1
0
Units of Measure
bits 0 - 7
A value of 0 for pounds, or 1 for kilograms
Decimal point position (places to the right of the decimal)
bits 8 - 15
A value from 0 to 5
1
1
Totalizer decimal point position (places to the right of the decimal)
bis 0 - 7
A value from 0 to 5
C2™, Second Generation Calibration
bits 8 - 15
Load Cell Count
1
2
Display Graduation size (‘count by): A value of 1,2,5,10,20,50,100,200, or 500
1
3