Hardy HIBSD Bench Scale
Page - 21 -
8.2. Linear Calibration
Liner calibration (multi-point calibration) can help to improve the accuracy of a scale over its full
range of weighing or to correct for binding issues.
NOTE:
To use linear calibration, the scale capacity must first be set by navigating to
R1 Cap
using
the method outlined in the PARAMETERS section.
•
After the scale capacity is set, navigate to
cal
and press
to confirm selection.
•
Press
once to select
liner
and
to confirm selection
•
Enter the password by pressing
,then
, then
to begin the calibration
process. The display will show
load 0
(no load on the scale).
•
Remove all the weight from the scale.
•
After the scale stabilizes and the zero Indicator comes on, press the
key to confirm.
Display will show
Load1.
•
Place the calibration weight (
1/3 of the capacity
) on the scale.
•
After the scale stabilizes and the zero Indicator comes on, press the
key to confirm.
Display will show
Load 2
•
Place the calibration weight (2/3 of the capacity) on the scale.
•
After the scale stabilizes and the zero Indicator comes on, press the
key to confirm.
Display will be shown
Load3
•
Place the calibration weight (full capacity) on the scale.
•
After the scale stabilizes and the zero Indicator comes on, press the
key to confirm.
Display will show
Pass
indicating the calibration process is complete. After self-check, the
Indicator will return to normal weighing mode.
Содержание BW Series
Страница 1: ...HIBSD Washdown Bench Scales Weighing Indicator Technical Manual ...
Страница 19: ...Hardy HIBSD Bench Scale Page 19 ...
Страница 22: ...Hardy HIBSD Bench Scale Page 22 10 DRAWING ...