BX176 portable single gas detector operation manual
15
At this time, press
or if there is no operation within one
minute, the detector will consider the current value to be the
calibration gas concentration value and go into calibrating mode.
Note
:
Adjusting range for calibration point please refer to
attached list 1.
4
)
Span calibration
In span calibration mode, the detector will show a
detecting value in current situation, see as below picture.
At this time, cover the gas sensing hole of the detector by
using calibration cap, open the gas valve, adjust the gas flow to be
120mL/min, when the detector detect the gas available for
calibration within 30s, it will start the analysis program for
calibrating automatically, and the display value will slowly go up
to calibration concentration which you already set, a few seconds
after then, the calibration will be completed and screen go to next
interface of alarm adjust automatically. Or else, if calibration fails,
screen will show “E000”.
Note
:
①
The gas flow is taking CO gas as an example,
forothergases calibration, please contact to the seller.
②
During span calibration, please do not press anybuttons,
otherwise the calibrating accuracy will be influenced or even the
detector can not work.