AT8100 Alcohol Tester operation manual
6
Connect the port of tester communication data with computer by the USB data cable, open
the computer and operate the coordinate data management software.
(
1
)
Turn on the tester, operate the data management software in the stand-by mode.
Notice: data upload must operate in the stand-by mode, do not operate the tester
during the process in case of mistakes of uploading.
(
2
)
Confirm the data backup to the computer, then turn off the tester.
8. Battery change
When the tester reminds low battery, or it can not work properly because of low power,
please change batteries in time.
9
.
Faults and solutions
Fault Description
Possible Reason
Solutions
Unable to turn on the tester
Low battery
Change batteries in time
Circuit fault
Contact with dealers or
manufacturer
Wrong Display
Expired sensor
Contact with manufacturer to
replace the battery
Over the specified
working conditions
Please be used in specified
conditions
Long time no calibration
Calibrate the tester
Wrong Time Display
Wrong time setup
Reset the time
Electromagnet
misoperation
No calibration for
pressure sensors
Calibrate pressure sensors again
10 Precaution
Please read the manual carefully and follow the instruction before your testing.
To ensure the real testing result, please wait 15 min to take the testing after drinking.
The calibration of the tester should not be longer than 6 months for ensuring the
accuracy of tester.
Please do not blow smoke to the tester directly, otherwise the tester will be damaged.
After storing for long time, the first test of tester may have errors.
Avoid using the tester when environmental temperature exceeds its specified range.
Avoid any high crash or strenuous vibration.
Any application or fault beyond this manual, please contact with manufacturer or
dealer for solutions.
11 Product accessories
Tester accessories include following:
1pc tester, 1pc printer
(
optional
)
, 1pc charger of printer
(
optional
)
,
1pc manual, 1pc inspection report, 1pc communication CD,
1pc data communication cable, 10pcs mouthpieces.