9
ORP MEASUREMENTS (HI 9125 only)
To perform ORP measurements, connect an optional ORP electrode
(see “Accessories” section) to the meter and turn it ON.
If necessary, enter the mV mode by pressing
RANGE
until the display
changes to mV.
Submerse the ORP electrode 3 cm (1¼”) into the sample to be tested
and wait a few minutes for the reading to stabilize.
Measurements within the ±699.9 mV range are displayed with
0.1 mV resolution, while outside this range the resolution automatically
switches to 1 mV.
For accurate ORP measurements, the surface of the electrode must be
clean and smooth. Pretreatment solutions are available to condition
the electrode and improve its response time (see “Accessories” section).
MEM & MR FUNCTIONS
The instrument allows the user to store the current measurement (pH
and temperature, or mV and temperature) into internal memory by
pressing the
MEM
key. The “
MEMORIZE
” tag lights up on display.
Stored values can be recalled by pressing
MR
: the display will show
the values and the “
RECALL MEMORIZED
” tag as long as the
button is pressed.