Liquid Crystal 
Display

CAL button

Operation

Preliminary Examination

Remove the meter from the packing material and examine it carefully to make sure that 

no damage has occurred during shipment. If noticeable damage is evident, contact your 

local Hanna Instruments Office.

Each meter is supplied with:

• CR2032 battery (1 pc.)

• Storage / Protection sleeve

• Instruction manual

• Quality Certificate

Note: Save all packing material until you are sure that the instrument functions 

correctly. All defective items must be returned in the original packaging with the 

supplied accessories.

Preparation

The probe is shipped dry. Before using the electrode, remove the protective cap. Then 

follow the calibration procedure.

• Turn the tester on by pressing ON/OFF button.
• Remove the protective cap and immerse the tip of the probe in the sample to 

be tested.

• Stir gently and wait for a stability tag to disappear.
• The electrode automatically compensates for temperature variations.
• The reading on display is directly expressed in: ppm.
• For best results, recalibrate periodically.
• After use rinse the probe with water.
• Always replace the protective cap after each use.

DiST 1

ppm

CAL

Specifications

TDS Conversion factor (CF)

0.5

Calibration 

Solutions

HI70032

(1382 ppm)                           

Calibration

automatic, single point (1382 ppm)

Temperature Compensation

automatic, 0.0 to 50.0 ºC (32.0 to 122.0 ºF)

Battery type 

CR2032 Li-Ion 3V (1 pc.)

Battery life

approximately 250 hours of continuous use

Environment

0 to 50 ºC (32 to 122 ºF);

RH 100% max

Weight
(without battery)

68 g (2.4 oz.)

ON/OFF button

Temperature 
Sensor (inside)

Probe

Range

0 to 2000 ppm

0.0 to 50.0 

˚

C (32.0 to 122.0 

˚

F)

Resolution

1 ppm 

0.1 

˚

C (0.1 

˚

F)

Accuracy 

(@25 °C/77 °F)

±2% F.S.
±0.5 °C (±1.0 °F)

All rights are reserved. Reproduction in whole or in part is prohibited 

without the written consent of the copyright owner.

Hanna Instruments reserves the right to modify the design, 

construction, or appearance of its products without advance notice.

• 1382 ppm calibration solution sachet (4 pcs.)

Operational Guide

To turn the meter ON and check the battery status

Press the ON/OFF button to turn the meter on. At start-up, all the LCD segments are 

displayed for 1 second, then the percent indication of the remaining battery life is 

displayed for another second. The meter then enters the normal measuring mode.

Note: Keeping the ON/OFF button pressed while turning the meter on will 

display all LCD segments as long as the button is pressed.

To enter calibration mode

Press the CAL button. “CAL“ message is displayed.

To enter setup mode

Remove the battery cover and press the Setup button located on the side of the battery.

Meter Setup

While in measurement mode, remove the battery cover. Press the Setup button 

located on the side of the battery in the battery compartment. The meter will enter 

in setup mode. Press the ON/OFF button to move through setup parameters. Change 

option by pressing CAL button. The default settings are: “Set t” measure unit - °C,  

“AOFF” - 8 min.

To select the temperature unit (°C/°F)

To select the temperature unit when “SET t” is displayed press the CAL button to 

change between °C or °F.

To select the Auto-Off time

To select the AUTO OFF when “AOFF” is displayed press the CAL button to change 

between 8 min, 60 min or --- (disabled).

To return to measurement mode

Press the ON/OFF button.

Setup 
Button

Recommended

Immersion 

Level

www.

.com

[email protected]

1.800.561.8187

Содержание DiST 1

Страница 1: ...INSTRUCTION MANUAL HI98301 TDS Tester DiST 1 ...

Страница 2: ...Liquid Crystal Display CAL button Operation DiST 1 ppm CAL ON OFF button Temperature Sensor inside Probe Recommended Immersion Level ...

Страница 3: ...ned in the original packaging with the supplied accessories Preparation The probe is shipped dry Before using the electrode remove the protective cap Then follow the calibration procedure Turn the tester on by pressing ON OFF button Remove the protective cap and immerse the tip of the probe in the sample to be tested Stir gently and wait for a stability tag to disappear The electrode automatically...

Страница 4: ...ompensation automatic 0 0 to 50 0 ºC 32 0 to 122 0 ºF Battery type CR2032 Li Ion 3V 1 pc Battery life approximately 250 hours of continuous use Environment 0 to 50 ºC 32 to 122 ºF RH 100 max Weight without battery 68 g 2 4 oz Range 0 to 2000 ppm 0 0 to 50 0 C 32 0 to 122 0 F Resolution 1 ppm 0 1 C 0 1 F Accuracy 25 C 77 F 2 F S 0 5 C 1 0 F ...

Страница 5: ...tery cover and press the Setup button located on the side of the battery Meter Setup While in measurement mode remove the battery cover Press the Setup button located on the side of the battery in the battery compartment The meter will enter in setup mode Press the ON OFF button to move through setup parameters Change option by pressing CAL button The default settings are Set t measure unit C AOFF...

Страница 6: ... message is displayed and meter returns to measurement mode Calibration Procedure Measurement and Calibration For better accuracy frequent calibration of the meter is recommended In addition the instrument must be recalibrated When high accuracy is required At least once a month Measurement Press the ON OFF button to turn the meter ON Place the probe to the desired solution The conductivity value ...

Страница 7: ...in accordance with local regulations The meter features a low battery indicator When the battery is running low under 10 the battery indicator will blink on the LCD When the battery is discharged dEAd bAtt will be displayed on the LCD for 2 seconds and the meter will turn off 0 0 Before using Hanna Instruments products make sure that they are entirely suitable for your specific application and for...

Страница 8: ...report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization RGA number from the Technical Service department and then send it with shipping costs prepaid When shipping any instrument m...

Страница 9: ...IST98301N 01 17 www com information itm com 1 800 561 8187 ...

Отзывы: