pH ELECTRODE MAINTENANCE

Preparation

Remove the electrode protective cap. Do not be alarmed if any salt deposits are
present: this is normal with electrodes and they will disappear when rinsed with
water.
During transport tiny bubbles of air may have formed inside the glass bulb. The
electrode cannot function properly under these conditions. These bubbles can be
removed by "shaking down" the electrode as you would do with a glass
thermometer.

Sensitive bulb

To avoid clogging problems and ensure a fast response, the bulb must be kept
moist at any time. Store the electrode with few drops of 

HI 70300 

storage

solution in the protective cap. NEVER USE DISTILLED OR DEIONIZED WATER FOR
STORING PURPOSES.

Periodic maintenance

Inspect electrode and cable. The cable used for the connection to the meter must
be intact and there must be no points of broken insulation. The connector must
be perfectly clean and dry. If any scratches or cracks are present, replace the
electrode.

Refilling the electrode

ALWAYS VERIFY THE ELECTROLYTE LEVEL.
If the electrolyte level is more than 2 cm below
the refilling hole, unscrew and remove the refill-
ing hole cap.
Add the 3.5M KCl electrolyte solution (

HI 7082

)

by using the supplied syringe. Tighten back the
cap. Wait at least 1 hour before using the
electrode.

pH VALUES AT VARIOUS TEMPERATURES

pH readings are strongly affected by temperature. The calibration buffer
solutions are affected by temperature changes to a lesser degree than normal
solutions.
For manual temperature calibration please refer to the following chart.

Note:

 To prevent damage to the electrode, remove the pH electrode from the

sample before turning the meter off. With the meter OFF, detach the
electrode from the meter and immerse it in the storage solution.

        TEMP                    pH VALUES

°C

°F

3.00

6.86

7.01

0
5

10
15
20
25
30
35
40
45
50
55
60
65
70

32
41
50
59
68
77
86
95

104
113
122
131
140
149
158

3.07
3.05
3.03
3.02
3.00
3.00
3.00
3.00
2.99
2.99
2.99
2.99
2.99
2.99
3.00

6.98
6.95
6.92
6.90
6.88
6.86
6.85
6.84
6.84
6.83
6.83
6.84
6.84
6.85
6.85

7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99

Содержание pH 209V

Страница 1: ...Instruction Manual w w w h a n n a i n s t c o m pH 209V Complete Laboratory Kit for pH ORP Measurements in Wine Must and Grape Juice ...

Страница 2: ...des pH 209 bench pH meter HI 1048B pH electrode HI 98501 electronic thermometer pH 3 and pH 7 buffer solutions 20 mL sachet each HI 700635 and HI 700636 specific cleaning solutions 20 mL each HI 7082 electrode refilling solution 30 mL bottle and syringe HI 76404 electrode holder 12 Vdc adapter and instruction manual Note Save all packing material until you are sure that the instrument functions co...

Страница 3: ...le with BNC and 1 m 3 3 cable Input Impedance 1012 ohm Power Supply 12 Vdc adapter included Environment 0 to 50 C 32 to 122 F RH max 95 non condensing Dimensions 240 x 182 x 74 mm 9 4x7 2x2 9 Weight 1 0 kg 2 2 lb meter only SPECIFICATIONS 1 Liquid Crystal Display LCD 2 pH slope adjustment knob 3 pH offset adjustment knob 4 pH range selection key 5 mV range selection key 6 C range selection key 7 T...

Страница 4: ...Vdc adapter into the power supply socket 9 on page 4 Note Make sure the main line is protected by a fuse Electrode connection For pH or ORP electrodes with internal reference connect the electrode to the BNC socket on the rear panel 10 on page 4 For electrodes with separate reference plug the BNC connector of the electrode to the BNC socket 10 on page 4 and the reference electrode to the proper so...

Страница 5: ...rence Switch the instrument on by pressing the ON OFF button Rinse and immerse the pH electrode in pH 7 01 buffer and shake briefly Wait a couple of minutes for the reading to stabilize Note the temperature of the buffer solution using the supplied HI 98501 thermometer e g 20 C Press the C key to select temperature setting then adjust the Temperature knob until the LCD shows the noted temperature ...

Страница 6: ...e the electrode Refilling the electrode ALWAYS VERIFY THE ELECTROLYTE LEVEL If the electrolyte level is more than 2 cm below the refilling hole unscrew and remove the refill ing hole cap Add the 3 5M KCl electrolyte solution HI 7082 by using the supplied syringe Tighten back the cap Wait at least 1 hour before using the electrode pH VALUES AT VARIOUS TEMPERATURES pH readings are strongly affected ...

Страница 7: ...trodes ORP PRETREATMENT SOLUTIONS HI 7091L Reducing pretreatment solution 500 mL bottle HI 7092L Oxidizing pretreatment solution 500 mL bottle Cleaning procedure After use clean the electrode by immersing it in the proper cleaning solution HI 700635 or HI 700636 for 5 minutes If case of slow and unreliable readings it is recommended to perform the following procedure for cleaning the junction 1 Sh...

Страница 8: ...g calibration of instruments ESD wrist straps should be worn to avoid possible damage to the electrode by electrostatic discharge To maintain the EMC performance of this equipment the recommended cables noted in the user s manual must be used Any variation introduced by the user to the supplied equipment may degrade the instrument s EMC performance To avoid electrical shock do not use this instrum...

Страница 9: ...2942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014 Fax 886 2 2739 2983 Thailand Tel 66 2619 0708 Fax 66 2619 0061 UnitedKingdom Tel 01525 850 855 Fax 01525 853 668 USA Tel 401 765 7500 Fax 401 765 7575 For e ma...

Отзывы: