13
12
• After confirmation, the selected param-
eter is displayed. The user can scroll
through the parameters by pressing CFM.
In order to directly set another param-
eter, press SETUP again and enter the
code or scroll to it by pressing CFM.
The following table lists the setup codes along with the descrip-
tion of the specific setup items, their valid values and whether
password is required to view that item (“PW” column):
Code
Valid Values
Default
PW
00 Factory ID
0 to 9999
0000
no
01 Process ID
0 to 9999 (models w/o RS485) 0000
no
0 to 99 (models with RS485) 00
no
02 Control enable/disable
0: C.M. disabled
0
no
1: C.M. enabled
11 Relay 1 mode
0: disabled
0
no
(M1)
1: ON-OFF high setpoint
2: ON-OFF low setpoint
3: PID, high setpoint
4: PID, low setpoint
for pH 5025YZ and mV 6025YZ only:
0: disabled
1: PID, high setpoint
2: PID, low setpoint
12 Relay 1 setpoint
0.00 to 14.00 pH
8.00 pH
no
(S1)
-2000 to 2000 mV
500 mV
13 Relay 1 hysteresis
0.00 to 14.00 pH
1 pH
no
(H1)
0 to 4000 mV
50 mV
14 Relay 1 deviation
0.50 to 14.00 pH
1 pH
no
(D1)
25 to 4000 mV
50 mV
15 Relay 1 reset time
0.1 to 999.9 minutes
999.9 mins
no
16 Relay 1 rate time
0.0 to 999.9 minutes
0.0 mins
no
21* Relay 2 mode (M2)
same as relay 1
0
no
* Available only in models with two relays
• Then confirm the displayed digit with
and move to the next one.
• When the whole password has been
inserted, press CFM to confirm it.
Note
The default password is set at “0000”.
• The LCD will display “SET” on the
upper part and “c.00” on the lower,
allowing the user to edit setup pa-
rameters (see table below).
• Enter the code of the parameter you want to set, using the
arrow keys as per the password procedure above (e.g. 41).
• Confirm the code by pressing CFM and the default or the
previously memorized value will be displayed with the first
digit blinking.
Note
When the password is not inserted or a wrong password is
confirmed, the display will only show the previously memo-
rized value, without blinking (read only mode). In this case,
the value cannot be set. Press LCD and start again.
• Enter the desired value using the arrow keys and then press
CFM.