16
PARAMETER SETUP MENU
ABSOLUTE EC RESOLUTION UNIT (HI763093 EC PROBE ONLY)
Option: Auto, Auto mS/cm, 1µS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm, 1 mS/cm
Press the
Modify
key to change the value. Use the
/
arrow key to select the value. Press
Select
to confirm or
ESC
to return to the menu without saving.
Absolute conductivity displays the conductivity measurement without temperature compensation. A
small letter “A” is added to the measurement unit to indicate no temperature compensation.
TDS RESOLUTION UNIT (HI763093 EC PROBE ONLY)
Option: Auto, Auto ppt, 1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt, 1 ppt
Press the
Modify
key to change the TDS resolution. Use the
/
arrow key to select the resolution.
Press
Select
to confirm or
ESC
to return to the menu without saving.
Auto allows the meter to choose the range that will optimize the measurement.
Auto ppt allows the meter to choose the range that will optimize the measurement, reading will only
be in ppt.
1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt or 1 ppt the meter will display the measurement in the selected
resolution.
DO CONCENTRATION UNIT (HI764103 DO PROBE ONLY)
Option: ppm or mg/L
Press the functional key to select the desired dissolved oxygen measurement unit.
7.3. PARAMETER COEFFICIENTS (EC & DO probes only)
From the
Parameter
setup menu, use the
/
keys to highlight Parameter coefficients and then
press
Select
.
Warning:
The previously saved logs for the connected probe will be updated with the selected
coefficients.