The Auto Power Off feature turns the instrument off after a set period (default 30 min) with no button
pressed to save battery life. To set another period or to disable this feature, see
SETUP
menu on
page 37. The instrument continues to monitor the inputs and memorize readings if the automatic
logging is enabled and started. To stop autologging press
StopLog
key or simply power off the
instrument by pressing the
ON/OFF
key.
The Auto Light Off backlight feature turns the backlight off after a set period (default 1 min) with
no buttons pressed. To set another period or to disable this feature, see
SETUP
menu on page 37.
MEASUREMENTS
Immerse the probe into the solution to be tested. The sleeve holes must be completely submerged. Tap
the probe repeatedly to remove any air bubbles that may be trapped inside the sleeve.
If needed, press
RANGE
repeatedly until the desired range (EC, Resistivity, TDS, Salinity) is selected
on the LCD.
Allow for the reading to stabilize. The main LCD line displays the measurement in the selected range,
while the temperature is displayed on the lower LCD line.
EC range
The conductivity range is from 0 to 400 mS/cm . The actual conductivity range (the uncompensated
conductivity) is up to 1000 mS/cm. The instrument will display conductivity readings up to 1000 mS/cm.
Note: The symbol in front of the temperature reading means that the temperature can be
entered by the user (Manual option selected in SETUP, or temperature out of range).
OPERA
TIONAL GUIDE